Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Radiat Prot Dosimetry ; 180(1-4): 37-41, 2018 Aug 01.
Artigo em Inglês | MEDLINE | ID: mdl-29788424

RESUMO

A long counter detector was manufactured by the Institute of Advanced Studies (IEAV) and was characterised in the neutron low scattering room at Brazilian National Ionising Radiation Metrology Laboratory (LNMRI/IRD) to deploy a secondary Standard for neutron fluence. The effective centre was measured experimentally with 252Cf+D2O, 252Cf, 241AmBe and 238PuBe neutron sources, having average energies from 0.55 to 4.16 MeV. The experimental arrangement and detector construction were carefully reproduced in Monte Carlo simulations, and the computational results were found to be in good agreement with those from experiment.


Assuntos
Amerício/normas , Berílio/normas , Califórnio/normas , Laboratórios/normas , Nêutrons , Plutônio/normas , Monitoramento de Radiação/normas , Proteção Radiológica/normas , Amerício/análise , Berílio/análise , Calibragem , Califórnio/análise , Método de Monte Carlo , Plutônio/análise , Doses de Radiação
2.
Radiat Prot Dosimetry ; 180(1-4): 56-61, 2018 Aug 01.
Artigo em Inglês | MEDLINE | ID: mdl-29474641

RESUMO

The standard thermal neutron flux unit, TNF2, in the Brazilian National Ionizing Radiation Metrology Laboratory was rebuilt. Fluence is still achieved by moderating of four 241Am-Be sources with 0.6 TBq each. The facility was again simulated and redesigned with graphite core and paraffin added graphite blocks surrounding it. Simulations using the MCNPX code on different geometric arrangements of moderator materials and neutron sources were performed. The resulting neutron fluence quality in terms of intensity, spectrum and cadmium ratio was evaluated. After this step, the system was assembled based on the results obtained from the simulations and measurements were performed with equipment existing in LNMRI/IRD and by simulated equipment. This work focuses on the characterization of a central chamber point and external points around the TNF2 in terms of neutron spectrum, fluence and ambient dose equivalent, H*(10). This system was validated with spectra measurements, fluence and H*(10) to ensure traceability.


Assuntos
Amerício/normas , Berílio/normas , Laboratórios/normas , Nêutrons , Monitoramento de Radiação/normas , Proteção Radiológica/normas , Amerício/análise , Berílio/análise , Calibragem , Método de Monte Carlo , Doses de Radiação
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA