IMPROVEMENTS IN THE THERMAL NEUTRON CALIBRATION UNIT, TNF2, AT LNMRI/IRD.
Astuto, A; Fernandes, S S; Patrão, K C S; Fonseca, E S; Pereira, W W; Lopes, R T.
Afiliação
  • Astuto A; Universidade Federal do Rio de Janeiro-COPPE/UFRJ. Av. Horácio Macedo, 2030, Bloco G/Sala 206-Centro de Tecnologia-Cidade Universitária, Ilha do Fundão, Rio de Janeiro, RJ, Brasil.
  • Fernandes SS; Universidade Federal do Rio de Janeiro-COPPE/UFRJ. Av. Horácio Macedo, 2030, Bloco G/Sala 206-Centro de Tecnologia-Cidade Universitária, Ilha do Fundão, Rio de Janeiro, RJ, Brasil.
  • Patrão KCS; Fundação Técnico Educacional Souza Marques-FTESM. Av. Ernani Cardoso, 335, Cascadura, Rio de Janeiro, RJ, Brasil.
  • Fonseca ES; Laboratório Nacional de Metrologia das Radiações Ionizantes LNMRI/IRD. Av. Salvador Allende, s/n, Barra da Tijuca, Rio de Janeiro, RJ, Brasil.
  • Pereira WW; Laboratório Nacional de Metrologia das Radiações Ionizantes LNMRI/IRD. Av. Salvador Allende, s/n, Barra da Tijuca, Rio de Janeiro, RJ, Brasil.
  • Lopes RT; Fundação Técnico Educacional Souza Marques-FTESM. Av. Ernani Cardoso, 335, Cascadura, Rio de Janeiro, RJ, Brasil.
Radiat Prot Dosimetry ; 180(1-4): 56-61, 2018 Aug 01.
Article em En | MEDLINE | ID: mdl-29474641
The standard thermal neutron flux unit, TNF2, in the Brazilian National Ionizing Radiation Metrology Laboratory was rebuilt. Fluence is still achieved by moderating of four 241Am-Be sources with 0.6 TBq each. The facility was again simulated and redesigned with graphite core and paraffin added graphite blocks surrounding it. Simulations using the MCNPX code on different geometric arrangements of moderator materials and neutron sources were performed. The resulting neutron fluence quality in terms of intensity, spectrum and cadmium ratio was evaluated. After this step, the system was assembled based on the results obtained from the simulations and measurements were performed with equipment existing in LNMRI/IRD and by simulated equipment. This work focuses on the characterization of a central chamber point and external points around the TNF2 in terms of neutron spectrum, fluence and ambient dose equivalent, H*(10). This system was validated with spectra measurements, fluence and H*(10) to ensure traceability.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Berílio / Monitoramento de Radiação / Amerício / Laboratórios / Nêutrons Tipo de estudo: Health_economic_evaluation / Prognostic_studies Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Brasil País de publicação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Berílio / Monitoramento de Radiação / Amerício / Laboratórios / Nêutrons Tipo de estudo: Health_economic_evaluation / Prognostic_studies Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Brasil País de publicação: Reino Unido