RESUMO

A custom-built controlled-rate freezing machine is described. It operates by automatically raising and lowering the sample carrier in a static column of liquid nitrogen vapour. It is controlled by a computer-assisted thermocouple feedback system that operates a stepper motor driving the sample carrier. The cooling protocol is divided into three phases: cooling from +5 to -5 C, initiation of ice nucleation, and cooling from -5 to -80 C. Experiments are described to validate the device over a range of different cooling rates. A freezing protocol is established to cool samples in plastic straws over a range of rates up to 80 C/min, with rapid and consistent absorption of the latent heat.


Assuntos
Criopreservação/instrumentação , Animais , Desenho de Equipamento , Congelamento , Masculino , Preservação do Sêmen/instrumentação , Suínos