Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Appl Radiat Isot ; 64(10-11): 1500-4, 2006.
Artigo em Inglês | MEDLINE | ID: mdl-16890447

RESUMO

During the last two years, a triple-to-double coincidence ratio (TDCR) system was assembled and adjusted at the CNEA-LMR, Argentina. The new counting system will add complementary capabilities to the absolute measurements section of the CNEA-LMR. This work describes its implementation and validation. Several checks and a set of beta-emitting standard solutions were used in order to perform the validation experiments. In preliminary measurements, a 3H LNHB solution with reference activity concentration of (119.7+/-0.9) kBq/g on 11 November 2003 was used. The CNEA-LMR TDCR counter gave, at the same reference date, an activity concentration of (120+/-1) kBq/g. Results and improvements are presented in detail. Concerning the asymmetry of the system, the quantum efficiency of the three photomultiplier tubes was studied for different operating conditions of the focusing voltage. The counter also includes an automatic system to change the efficiency by defocusing the photomultipliers and on the other hand, it was coupled to a HPGe detector to also measure beta-gamma coincidences.


Assuntos
Algoritmos , Modelos Químicos , Radioisótopos/análise , Radioisótopos/normas , Contagem de Cintilação/instrumentação , Contagem de Cintilação/métodos , Argentina , Simulação por Computador , Doses de Radiação , Radioisótopos/química , Padrões de Referência , Valores de Referência , Contagem de Cintilação/normas
2.
Appl Radiat Isot ; 63(2): 229-33, 2005 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-15896971

RESUMO

The activity concentration of a (238)Pu solution was measured by the determined solid angle method employing a novel dual diaphragm-detector assembly, which has been previously described. Due to the special requirements of the detector, a new type of source holder was developed, which consisted of sandwiching the radioisotope between two organic films called VYNS. It was experimentally demonstrated that the VYNS films do not absorb alpha particles, but reduce their energy by an average of 22 keV.A mean activity concentration for (238)Pu of 359.10+/-0.8 kBq/g was measured.


Assuntos
Algoritmos , Plutônio/análise , Radiometria/instrumentação , Manejo de Espécimes/instrumentação , Partículas alfa , Desenho Assistido por Computador , Desenho de Equipamento , Análise de Falha de Equipamento , Doses de Radiação , Radiometria/métodos , Soluções , Manejo de Espécimes/métodos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA