Determination of the activity concentration of a 238 Pu solution by the defined solid angle method utilizing a novel dual diaphragm-detector assembly.
Appl Radiat Isot
; 63(2): 229-33, 2005 Aug.
Article
em En
| MEDLINE
| ID: mdl-15896971
The activity concentration of a (238)Pu solution was measured by the determined solid angle method employing a novel dual diaphragm-detector assembly, which has been previously described. Due to the special requirements of the detector, a new type of source holder was developed, which consisted of sandwiching the radioisotope between two organic films called VYNS. It was experimentally demonstrated that the VYNS films do not absorb alpha particles, but reduce their energy by an average of 22 keV.A mean activity concentration for (238)Pu of 359.10+/-0.8 kBq/g was measured.