1.
Radiother Oncol
; 90(1): 56-9, 2009 Jan.
Artículo
en Inglés
| MEDLINE
| ID: mdl-18417232
RESUMEN
Surgical clips were investigated for partial breast image-guided radiotherapy (IGRT). Small titanium clips were insufficiently well visualised. Medium tantalum clips were best for megavoltage IGRT and small tantalum clips were best for floor mounted kilovoltage IGRT (ExacTrac). Both small tantalum and medium titanium clips were suitable for isocentric kilovoltage IGRT.