An anthropomorphic phantom study of visualisation of surgical clips for partial breast irradiation (PBI) setup verification.
Thomas, Carys W; Nichol, Alan M; Park, Julie E; Hui, Jason F; Giddings, Alison A; Grahame, Sheri; Otto, Karl.
Afiliación
  • Thomas CW; Department of Radiation Oncology, Vancouver Centre, Vancouver, BC, Canada.
Radiother Oncol ; 90(1): 56-9, 2009 Jan.
Article en En | MEDLINE | ID: mdl-18417232
Surgical clips were investigated for partial breast image-guided radiotherapy (IGRT). Small titanium clips were insufficiently well visualised. Medium tantalum clips were best for megavoltage IGRT and small tantalum clips were best for floor mounted kilovoltage IGRT (ExacTrac). Both small tantalum and medium titanium clips were suitable for isocentric kilovoltage IGRT.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Instrumentos Quirúrgicos / Neoplasias de la Mama / Fantasmas de Imagen Tipo de estudio: Risk_factors_studies Límite: Humans Idioma: En Revista: Radiother Oncol Año: 2009 Tipo del documento: Article País de afiliación: Canadá Pais de publicación: Irlanda

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Instrumentos Quirúrgicos / Neoplasias de la Mama / Fantasmas de Imagen Tipo de estudio: Risk_factors_studies Límite: Humans Idioma: En Revista: Radiother Oncol Año: 2009 Tipo del documento: Article País de afiliación: Canadá Pais de publicación: Irlanda