Dose rate constants for 125I, 103Pd, 192Ir and 169Yb brachytherapy sources: an EGS4 Monte Carlo study.
Phys Med Biol
; 43(6): 1557-66, 1998 Jun.
Article
em En
| MEDLINE
| ID: mdl-9651025
An exhaustive revision of dosimetry data for 192Ir, 125I, 103Pd and 169Yb brachytherapy sources has been performed by means of the EGS4 simulation system. The DLC-136/PHOTX cross section library, water molecular form factors, bound Compton scattering and Doppler broadening of the Compton-scattered photon energy were considered in the calculations. The absorbed dose rate per unit contained activity in a medium at 1 cm in water and air-kerma strength per unit contained activity for each seed model were calculated, allowing the dose rate constant (DRC) A to be estimated. The influence of the calibration procedure on source strength for low-energy brachytherapy seeds is discussed. Conversion factors for 125I and 103Pd seeds to obtain the dose rate in liquid water from the dose rate measured in a solid water phantom with a detector calibrated for dose to water were calculated. A theoretical estimate of the DRC for a 103Pd model 200 seed equal to 0.669 +/- 0.002 cGy h(-1) U(-1) is obtained. Comparison of obtained DRCs with measured and calculated published results shows agreement within 1.5% for 192Ir, 169Yb and 125I sources.