The permeability of tissue expanders to bacteria: an experimental study.
Liang, M D; Narayanan, K; Ravilochan, K; Roche, K.
Afiliação
  • Liang MD; Division of Plastic Surgery, University of Pittsburgh Medical Center, Pa.
Plast Reconstr Surg ; 92(7): 1294-7, 1993 Dec.
Article em En | MEDLINE | ID: mdl-8248404
Infection is an occasional but serious complication of tissue expansion. Bacteria in many cases are found both inside and outside the expander. The mechanism of how they spread inside and out is not fully understood. In the current study, we have assessed both the expander membranes (n = 28) and the port (n = 42) in terms of leakage and transmission of Staphylococcus aureus in normal expansion (100 percent) and overexpansion (400 percent) in an in vitro model. Our results indicate that the membrane was impermeable to S. aureus, whereas the bacteria were able to migrate freely through the puncture sites in the ports.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Staphylococcus aureus / Dispositivos para Expansão de Tecidos / Expansão de Tecido / Contaminação de Equipamentos Idioma: En Revista: Plast Reconstr Surg Ano de publicação: 1993 Tipo de documento: Article País de publicação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Staphylococcus aureus / Dispositivos para Expansão de Tecidos / Expansão de Tecido / Contaminação de Equipamentos Idioma: En Revista: Plast Reconstr Surg Ano de publicação: 1993 Tipo de documento: Article País de publicação: Estados Unidos