Repeatability and intra and interrater reproducibility of the DIPA-S eHealth© capture and analysis system for clinical assessment of scoliosis.
Navarro, Isis Juliene Rodrigues Leite; Silva, Marcelle Guimarães; Dos Santos, Ingrid Cláudia Pereira; Candotti, Cláudia Tarragô.
Afiliação
  • Navarro IJRL; Universidade Federal do Rio Grande do Sul - Grupo Biomec, Rua Felizardo, 750, bairro Jardim Botânico, Porto Alegre, Rio Grande do Sul, Brasil. isisjlnavarro@gmail.com.
  • Silva MG; Universidade Federal do Rio Grande do Sul - Grupo Biomec, Rua Felizardo, 750, bairro Jardim Botânico, Porto Alegre, Rio Grande do Sul, Brasil.
  • Dos Santos ICP; Universidade Federal do Rio Grande do Sul - Grupo Biomec, Rua Felizardo, 750, bairro Jardim Botânico, Porto Alegre, Rio Grande do Sul, Brasil.
  • Candotti CT; Universidade Federal do Rio Grande do Sul - Grupo Biomec, Rua Felizardo, 750, bairro Jardim Botânico, Porto Alegre, Rio Grande do Sul, Brasil.
Eur Spine J ; 33(10): 3798-3805, 2024 Oct.
Article em En | MEDLINE | ID: mdl-39014079
ABSTRACT

PURPOSE:

To assess the repeatability, intra and interrater reproducibility of the DIPA-S eHealth© system for capturing and measuring clinical variables of scoliosis, including frontal trunk imbalance (FTI), sagittal trunk imbalance (STI), and angle of trunk rotation (ATR).

METHODS:

Patients were photographed using the DIPA-S eHealth Capture© mobile application by family members, physiotherapists, or surgeons. Three photos were taken in each position standing in the frontal and sagittal planes and in the axial plane in forward bending position of the trunk. The photos were analyzed by three independent evaluators using the DIPA-S eHealth Analysis© software. For repeatability, each photo was analyzed twice consecutively by the three evaluators. For intrarater reproducibility, only one evaluator reanalyzed the first photo from each plane with a 5-day interval. For interrater reproducibility, the three evaluators analyzed the first photo from each plane. The intraclass correlation coefficient (ICC), standard error of measurement (SEM), and minimal detectable change (MDC) were used (p < 0.05).

RESULTS:

The sample comprised 262 images (n = 30). Repeatability was excellent for all three evaluators in all three planes (ICC 0.94 to 1.00). Intrarater reproducibility was excellent in all three planes (ICC 0.88 to 0.99). Interrater analysis was excellent for the frontal and axial planes (ICC 0.98 and 0.93), respectively. However, it was weak in the sagittal plane ICC 0.32. The SEM ranged from 0.02 to 17.6 and MDC from 0.03 to 34.6.

CONCLUSION:

The DIPA-S eHealth© Capture and Analysis system demonstrates reproducibility for use in the clinical assessment of scoliosis through teleconsultations.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Escoliose / Telemedicina Limite: Adolescent / Child / Female / Humans / Male Idioma: En Revista: Eur Spine J Assunto da revista: ORTOPEDIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Brasil País de publicação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Escoliose / Telemedicina Limite: Adolescent / Child / Female / Humans / Male Idioma: En Revista: Eur Spine J Assunto da revista: ORTOPEDIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Brasil País de publicação: Alemanha