Vertical root fracture diagnosis in teeth with metallic posts: Impact of metal artifact reduction and sharpening filters.
Ruiz, Débora Costa; Rosado, Lucas P Lopes; Fontenele, Rocharles Cavalcante; Farias-Gomes, Amanda; Freitas, Deborah Queiroz.
Afiliação
  • Ruiz DC; Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba, SP, Brazil.
  • Rosado LPL; Department of Dentistry, University Center of Espírito Santo, Colatina, ES, Brazil.
  • Fontenele RC; OMFS IMPATH Research Group, Department of Imaging and Pathology, Faculty of Medicine, KU Leuven, Leuven, Belgium.
  • Farias-Gomes A; Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba, SP, Brazil.
  • Freitas DQ; Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba, SP, Brazil.
Imaging Sci Dent ; 54(2): 139-145, 2024 Jun.
Article em En | MEDLINE | ID: mdl-38948185
ABSTRACT

Purpose:

This study examined the influence of a metal artifact reduction (MAR) tool, sharpening filters, and their combination on the diagnosis of vertical root fracture (VRF) in teeth with metallic posts using cone-beam computed tomography (CBCT). Materials and

Methods:

Twenty single-rooted human premolars - 9 with VRF and 11 without - were individually placed in a human mandible. A metallic post composed of a cobalt-chromium alloy was inserted into the root canal of each tooth. CBCT scans were then acquired under the following parameters 8 mA, a 5×5 cm field of view, a voxel size of 0.085 mm, 90 kVp, and with MAR either enabled or disabled. Five oral and maxillofacial radiologists independently evaluated the CBCT exams under each MAR mode and across 3 sharpening filter conditions no filter, Sharpen 1×, and Sharpen 2×. The diagnostic performance was quantified by the area under the receiver operating characteristic curve (AUC), sensitivity, and specificity. These metrics were compared using 2-way analysis of variance with a significance level of α=5%. Intra- and inter-examiner agreement were assessed using the weighted kappa test.

Results:

Neither MAR nor the application of sharpening filters significantly impacted AUC or specificity (P>0.05). However, sensitivity increased when MAR was combined with Sharpen 1× and Sharpen 2× (P=0.015). The intra-examiner agreement ranged from fair to substantial (0.34-0.66), while the inter-examiner agreement ranged from fair to moderate (0.27-0.41).

Conclusion:

MAR in conjunction with sharpening filters improved VRF detection; therefore, their combined use is recommended in cases of suspected VRF.