Your browser doesn't support javascript.
loading
Chitosan/genipin modified electrode for voltammetric determination of interleukin-6 as a biomarker of sepsis.
de Matos Morawski, Franciele; Dias, Greicy Brisa Malaquias; Sousa, Kelline Alaide Pereira; Formiga, Rodrigo; Spiller, Fernando; Parize, Alexandre Luis; Báfica, André; Jost, Cristiane Luisa.
Afiliação
  • de Matos Morawski F; ampere - Laboratório de Plataformas Eletroquímicas, Universidade Federal de Santa Catarina, Departamento de Química, CEP 88040-900 Florianópolis, SC, Brazil.
  • Dias GBM; Laboratory of Immunobiology, Universidade Federal de Santa Catarina, Departamento de Microbiologia, Imunologia e Parasitologia, CEP 88040-900, Florianópolis, SC, Brazil.
  • Sousa KAP; ampere - Laboratório de Plataformas Eletroquímicas, Universidade Federal de Santa Catarina, Departamento de Química, CEP 88040-900 Florianópolis, SC, Brazil.
  • Formiga R; Laboratory of Immunobiology, Universidade Federal de Santa Catarina, Departamento de Microbiologia, Imunologia e Parasitologia, CEP 88040-900, Florianópolis, SC, Brazil.
  • Spiller F; Laboratory of Immunobiology, Universidade Federal de Santa Catarina, Departamento de Microbiologia, Imunologia e Parasitologia, CEP 88040-900, Florianópolis, SC, Brazil.
  • Parize AL; ampere - Laboratório de Plataformas Eletroquímicas, Universidade Federal de Santa Catarina, Departamento de Química, CEP 88040-900 Florianópolis, SC, Brazil.
  • Báfica A; Laboratory of Immunobiology, Universidade Federal de Santa Catarina, Departamento de Microbiologia, Imunologia e Parasitologia, CEP 88040-900, Florianópolis, SC, Brazil.
  • Jost CL; ampere - Laboratório de Plataformas Eletroquímicas, Universidade Federal de Santa Catarina, Departamento de Química, CEP 88040-900 Florianópolis, SC, Brazil. Electronic address: franciele.morawski@ufsc.br.
Int J Biol Macromol ; 224: 1450-1459, 2023 Jan 01.
Article em En | MEDLINE | ID: mdl-36441080
Ultrasensitive electroanalytical monitoring of interleukin-6 levels in serum samples has emerged as a valuable tool for the early diagnosis of inflammatory diseases. Despite its advantages, there is a lack of strategies for the label-free voltammetric determination of cytokines. Here, a novel chitosan/genipin modified fluorine tin oxide electrode was developed providing an in-situ hydrogel formation (FTO/CSG). This platform was applied for the detection of interleukin-6, a major pro-inflammatory cytokine. Transmission electron microscopy (TEM), cyclic voltammetry (CV), and electrochemical impedance spectroscopy (EIS) indicated genipin serves as an efficient green cross-linker to build the immunosensing platform (FTO/CSG/anti-IL-6). EIS showed an increase in charge transfer resistance from 326 to 1360 kΩ after the immobilization of anti-IL-6 antibodies. By square wave voltammetry, this method achieved a detection limit of 0.03 pg mL-1 with a wide linear range of 0.05-1000 pg mL-1. Additionally, it displayed a high selectivity index when tested in the presence of three inflammatory cytokines as interfering proteins: IL-12, IL-1ß, and TNF-α. The sample matrix effect showed a peak current variation lower than 5 %. The novel method was applied for the quantification of IL-6 in serum samples of septic mice. No statistical differences were observed between the standard ELISA and the proposed method using a confidence level of 95 %.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas Biossensoriais / Sepse / Quitosana Tipo de estudo: Screening_studies Limite: Animals Idioma: En Revista: Int J Biol Macromol Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Brasil País de publicação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas Biossensoriais / Sepse / Quitosana Tipo de estudo: Screening_studies Limite: Animals Idioma: En Revista: Int J Biol Macromol Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Brasil País de publicação: Holanda