Automatic exposure compensation of digital radiographic technologies does not affect alveolar bone-level measurement.
Gaêta-Araujo, Hugo; Oliveira-Santos, Nicolly; de Oliveira Reis, Larissa; Nascimento, Eduarda Helena Leandro; Oliveira-Santos, Christiano.
Afiliação
  • Gaêta-Araujo H; Oral Radiology Section, School of Dentistry, Federal University of Alfenas (UNIFAL-MG), Rua Gabriel Monteiro da Silva, 700, Alfenas, MG, 37130-000, Brazil. hugogaeta@hotmail.com.
  • Oliveira-Santos N; Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas (UNICAMP), Piracicaba, SP, Brazil.
  • de Oliveira Reis L; Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas (UNICAMP), Piracicaba, SP, Brazil.
  • Nascimento EHL; Department of Dentistry, Federal University of Juiz de Fora (UFJF-GV), Governador Valadares, MG, Brazil.
  • Oliveira-Santos C; Department of Dentistry, Division of Oral Radiology, Odontomed Imagem - Medical and Dental Services, Recife, PE, Brazil.
Oral Radiol ; 39(1): 53-58, 2023 01.
Article em En | MEDLINE | ID: mdl-35218461
OBJECTIVE: To evaluate whether the automatic exposure compensation in the presence of high-density materials can affect the measurement of alveolar bone level. METHODS: Thirty regions of seven dry skulls and six mandibles were radiographed with and without a high-density material, using two digital radiographic technologies: photostimulable phosphor plate (PSP, Digora Optime) and sensor (CMOS, Digora Toto), totaling 120 images. The distances from the cement-enamel junction to the alveolar bone crest were measured using cone-beam computed tomography (CBCT) images to represent the reference standard. The same measurements of alveolar bone level and the average of the pixel values of the image were evaluated on the radiographs. Paired t test compared the average pixel values and alveolar bone-level measurements between images with and without high-density material. One-way analysis of variance compared the difference between radiographic and CBCT measurements (α = 0.05). RESULTS: The high-density material reduced the pixel values in PSP (p = 0.002) and CMOS (p < 0.001) technologies, demonstrating the AEC functioning in both technologies. There was no difference in bone-level measurements between the images without and with the high-density material for both technologies (p ≥ 0.091), or between the tomographic and radiographic measurements (p ≥ 0.319). CONCLUSION: In the presence of high-density material, the automatic exposure compensation reduces the average pixel values of the images (i.e., images get darker), but does not influence the radiographic measurements of alveolar bone level.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiografia Dentária Digital / Processo Alveolar Idioma: En Revista: Oral Radiol Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Brasil País de publicação: Japão