Determination of chromium(III) picolinate in dietary supplements by flow injection - electrospray ionization - tandem mass spectrometry, using cobalt(II) picolinate as internal standard.
Arroyo Negrete, Missael Antonio; Wrobel, Kazimierz; Yanez Barrientos, Eunice; Corrales Escobosa, Alma Rosa; Enciso Donis, Israel; Wrobel, Katarzyna.
Afiliação
  • Arroyo Negrete MA; Chemistry Department, Division of Natural and Exact Sciences, University of Guanajuato, L. de Retana 5, 36000 Guanajuato, Mexico.
  • Wrobel K; Chemistry Department, Division of Natural and Exact Sciences, University of Guanajuato, L. de Retana 5, 36000 Guanajuato, Mexico.
  • Yanez Barrientos E; Chemistry Department, Division of Natural and Exact Sciences, University of Guanajuato, L. de Retana 5, 36000 Guanajuato, Mexico.
  • Corrales Escobosa AR; Chemistry Department, Division of Natural and Exact Sciences, University of Guanajuato, L. de Retana 5, 36000 Guanajuato, Mexico.
  • Enciso Donis I; Chemistry Department, Division of Natural and Exact Sciences, University of Guanajuato, L. de Retana 5, 36000 Guanajuato, Mexico.
  • Wrobel K; Chemistry Department, Division of Natural and Exact Sciences, University of Guanajuato, L. de Retana 5, 36000 Guanajuato, Mexico. Electronic address: katarzyn@ugto.mx.
Talanta ; 240: 123161, 2022 Apr 01.
Article em En | MEDLINE | ID: mdl-34953383
In this work, a principle of flow injection analysis (FIA) was applied for sample introduction to an electrospray ionization - ion trap mass spectrometer (ESI-ITMS) with the aim to quantify chromium(III) picolinate (CrPic3) in commercial supplements by multiple reaction monitoring, and using cobalt(II) picolinate as internal standard (IS). FIA system was operated with ammonium formate 10 mmol L-1 in methanol-water (1:1, v/v) as a carrier solution at a flow rate 200 µL min-1; 100 µL injections were performed in 2-min intervals. Setting ion transitions m/z 419 â†’ 270 and 304 â†’ 260 for the analyte and IS, respectively, and 100 ms integration time, the method detection and quantification limits 12 ng g-1 and 40 ng g-1 of Cr (as CrPic3) in the air-dried powder. Acetonitrile extracts of the real-world samples presented varying from sample-to-sample chemical composition and IS efficiently compensated for ionization interferences. Mean results from triplicate analysis of four different supplements were obtained with relative standard deviation 0.1-4.0%, indicating acceptable precision. Trueness of the proposed FIA-ESI-ITMS/MS procedure was demonstrated by 95.8-108% percentage recoveries attained in the analysis of the CrPic3-spiked samples. For comparative purposes, total Cr was determined by ICP-MS. The quantitative results obtained indicate the necessity of analytical control of Cr(III) supplements commercially available and demonstrate that the proposed FIA-ESI-ITMS/MS procedure is well-suited for the determination of CrPic3 in such products.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas por Ionização por Electrospray / Espectrometria de Massas em Tandem Idioma: En Revista: Talanta Ano de publicação: 2022 Tipo de documento: Article País de afiliação: México País de publicação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas por Ionização por Electrospray / Espectrometria de Massas em Tandem Idioma: En Revista: Talanta Ano de publicação: 2022 Tipo de documento: Article País de afiliação: México País de publicação: Holanda