Your browser doesn't support javascript.
loading
Assessment of the use of boron-doped diamond electrode for highly sensitive voltammetric determination of the azo-dye carmoisine E-122 in food and environmental matrices.
Micheletti, Lais; Coldibeli, Bruna; Salamanca-Neto, Carlos Alberto Rossi; Almeida, Lucio Cesar; Sartori, Elen Romão.
Afiliação
  • Micheletti L; Laboratório de Eletroanalítica e Sensores, Departamento de Química, Centro de Ciências Exatas, Universidade Estadual de Londrina, CEP 86057-970, Londrina - PR, Brazil.
  • Coldibeli B; Laboratório de Eletroanalítica e Sensores, Departamento de Química, Centro de Ciências Exatas, Universidade Estadual de Londrina, CEP 86057-970, Londrina - PR, Brazil.
  • Salamanca-Neto CAR; Laboratório de Eletroanalítica e Sensores, Departamento de Química, Centro de Ciências Exatas, Universidade Estadual de Londrina, CEP 86057-970, Londrina - PR, Brazil. Electronic address: carlos.salamanca@uel.br.
  • Almeida LC; Laboratório de Eletroquímica Ambiental, Departamento de Química, Centro de Ciências Exatas, Universidade Estadual de Londrina, CEP 86057-970, Londrina - PR, Brazil.
  • Sartori ER; Laboratório de Eletroanalítica e Sensores, Departamento de Química, Centro de Ciências Exatas, Universidade Estadual de Londrina, CEP 86057-970, Londrina - PR, Brazil. Electronic address: elensartori@uel.br.
Talanta ; 220: 121417, 2020 Dec 01.
Article em En | MEDLINE | ID: mdl-32928428
A new protocol for the analysis of the azo-dye carmoisine (CMS) is presented by coupling differential pulse voltammetry (DPV) with a cathodically pretreated boron-doped diamond electrode (CPT-BDDE), in phosphate buffer solution (pH 2.0). The CMS presented diffusion-controlled oxidation and reduction peaks at +0.88 and -0.15 V vs Ag/AgCl, respectively. The effect of the pretreatment conditions, pH, and supporting electrolytes were evaluated to the voltammetric determination of CMS. Under optimized conditions, the differential pulse voltammetric signals for CMS were linear over the concentration range of 0.059-1.31 µmol L-1 and 0.010-0.079 µmol L-1 with limits of detection of 7.0 and 3.0 nmol L-1, for the anodic and cathodic processes respectively. The method was precise for CMS determination (RSD < 5.0%) and selective against other dyes. The developed protocol was successfully applied in the analysis of CMS in surface water and foodstuffs with accurate results in comparison with those obtained using a validated spectrophotometric method.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Boro / Diamante Tipo de estudo: Diagnostic_studies Idioma: En Revista: Talanta Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Brasil País de publicação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Boro / Diamante Tipo de estudo: Diagnostic_studies Idioma: En Revista: Talanta Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Brasil País de publicação: Holanda