Intra and inter-raters reliability and agreement of stimulus electrodiagnostic tests with two different electrodes in sedated critically-ill patients.
de Araujo, Amaro Eduardo Tavares; Silva, Paulo Eugênio; Carvalho, Karina Livinode de; Fachin-Martins, Emerson; Babault, Nicolas; Durigan, João Luiz Quagliotti.
Afiliação
  • de Araujo AET; Physical Therapy Division, Hospital de Base do Distrito Federal. St. Médico Hospitalar Sul - Asa Sul , Brasília, Federal District, Brazil.
  • Silva PE; Physical Therapy Division, Hospital de Base do Distrito Federal. St. Médico Hospitalar Sul - Asa Sul , Brasília, Federal District, Brazil.
  • Carvalho KL; Physical Therapy Division, Hospital de Base do Distrito Federal. St. Médico Hospitalar Sul - Asa Sul , Brasília, Federal District, Brazil.
  • Fachin-Martins E; College of Ceilândia, Centro Metropolitano, University of Brasília , Brasília, Federal District, Brazil.
  • Babault N; Centre d'Expertise de la Performance G. Cometti, U1093-INSERM, CAPS, Faculté des Siences du Sport, Université de Bourgogne-Franche-Comté , Dijon, France.
  • Durigan JLQ; College of Ceilândia, Centro Metropolitano, University of Brasília , Brasília, Federal District, Brazil.
Physiother Theory Pract ; 36(12): 1447-1456, 2020 Dec.
Article em En | MEDLINE | ID: mdl-30739542
Objective: The aim of the present study was to verify the intra- and inter-rater reliability and agreement of the stimulus electrodiagnostic test (SET) measurements obtained by pen and square electrodes in the vastus lateralis and tibialis anterior muscles. Design: An intra- and inter-rater reliability and agreement study was performed for the SET by two independent raters. Two different sizes of cathode electrodes (1 cm2 and 25 cm2) and two muscles were assessed (tibialis anterior and vastus lateralis). Results: Chronaxie did not change according to the different electrodes. A high intra-rater reliability (0.72 ≤ r ≤ 0.88) was detected independently of the electrode and muscle assessed. Moreover, moderate and almost perfect agreements (0.51 ≤ Kappa ≤ 1.00) were detected on intra-rater assessment. Similar correlations (0.74 ≤ r ≤ 0.79) were found for intra-rater reliability. However, dissimilar inter-rater agreement was detected: Kappa ≤ 0.40 for tibialis anterior and Kappa = 1.00 for vastus lateralis. Conclusion: The SET presented high reliability and moderate agreement in intra-rater evaluations. A fair agreement was found in the inter-rater assessment of the tibialis anterior. Evaluations performed with different electrode sizes did not influence the results. Therefore, the SET should be performed by a unique rater in test retest situations.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Polineuropatias / Estado Terminal / Músculo Esquelético / Eletrodos / Eletrodiagnóstico Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Physiother Theory Pract Assunto da revista: MEDICINA FISICA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Brasil País de publicação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Polineuropatias / Estado Terminal / Músculo Esquelético / Eletrodos / Eletrodiagnóstico Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Physiother Theory Pract Assunto da revista: MEDICINA FISICA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Brasil País de publicação: Reino Unido