Tolerability and Functionality of a Wireless 24-Hour Ocular Telemetry Sensor in African American Glaucoma Patients.
Marando, Catherine M; Mansouri, Kaweh; Kahook, Malik Y; Seibold, Leonard K.
Afiliação
  • Marando CM; Department of Ophthalmology, University of Colorado Eye Center, Aurora, CO.
  • Mansouri K; Albert Einstein College of Medicine, Bronx, NY.
  • Kahook MY; Department of Ophthalmology, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, MA.
  • Seibold LK; Department of Ophthalmology, University of Colorado Eye Center, Aurora, CO.
J Glaucoma ; 28(2): 119-124, 2019 02.
Article em En | MEDLINE | ID: mdl-30688840
PURPOSE: The purpose of the study was to evaluate the tolerability and functionality of a wireless ocular telemetry sensor in African American patients with glaucoma. MATERIALS AND METHODS: In this prospective, observational cohort study, 20 African American patients with primary open angle glaucoma (POAG) were evaluated at the University of Colorado Eye Center. Before lens placement, patients recorded ocular comfort and underwent a baseline eye exam. Following the exam, patients were fitted with a SENSIMED Triggerfish contact lens sensor and data recording device. Patients were sent home and instructed to record their activities in a journal and return in 24 hours. Repeat exams were performed at various time points in clinic before and after lens removal. RESULTS: All 20 patients retained the lens for the 24-hour study period. The patient reported comfort was excellent, with a nadir of mean recorded comfort of 7.05/10. Significant clinical changes were noted in lid/conjunctival erythema, BCVA, refraction, and pachymetry over the course of lens wear. The majority of these changes were improved or resolved by 1 hour after lens removal. Voltage output was significantly greater nocturnally than diurnally (184.79 mV and 71.48 mV, respectively; P<0.0001). There was no significant change in signal variability or slope over the entire duration of the sleep/wake period based on sleep. CONCLUSIONS: The wireless ocular sensor is well tolerated over a 24-hour period in African American patients with POAG despite transient changes in visual acuity and conjunctival erythema. Clinically usable 24-hour profiles were generated for all patients, with voltage output increasing significantly during periods of sleep.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tonometria Ocular / Negro ou Afro-Americano / Glaucoma de Ângulo Aberto / Tecnologia de Sensoriamento Remoto / Pressão Intraocular Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Glaucoma Assunto da revista: OFTALMOLOGIA Ano de publicação: 2019 Tipo de documento: Article País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tonometria Ocular / Negro ou Afro-Americano / Glaucoma de Ângulo Aberto / Tecnologia de Sensoriamento Remoto / Pressão Intraocular Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Glaucoma Assunto da revista: OFTALMOLOGIA Ano de publicação: 2019 Tipo de documento: Article País de publicação: Estados Unidos