Your browser doesn't support javascript.
loading
Metrological Validation of a Measurement Procedure for the Characterization of a Biological Ultrasound Tissue-Mimicking Material.
Santos, Taynara Q; Alvarenga, André V; Oliveira, Débora P; Costa-Felix, Rodrigo P B.
Afiliação
  • Santos TQ; Laboratory of Ultrasound, Directory of Scientific and Industrial Metrology, National Institute of Metrology, Quality and Technology, Rio de Janeiro, Brazil.
  • Alvarenga AV; Laboratory of Ultrasound, Directory of Scientific and Industrial Metrology, National Institute of Metrology, Quality and Technology, Rio de Janeiro, Brazil. Electronic address: avalvarenga@inmetro.gov.br.
  • Oliveira DP; Laboratory of Ultrasound, Directory of Scientific and Industrial Metrology, National Institute of Metrology, Quality and Technology, Rio de Janeiro, Brazil.
  • Costa-Felix RP; Laboratory of Ultrasound, Directory of Scientific and Industrial Metrology, National Institute of Metrology, Quality and Technology, Rio de Janeiro, Brazil.
Ultrasound Med Biol ; 43(1): 323-331, 2017 01.
Article em En | MEDLINE | ID: mdl-27756496
The speed of sound and attenuation are important properties for characterizing reference materials such as biological phantoms used in ultrasound applications. There are many publications on the manufacture of ultrasonic phantoms and the characterization of their properties. However, few studies have applied the principles of metrology, such as the expression of the uncertainty of measurement. The objective of this study is to validate a method for characterizing the speed of sound and the attenuation coefficient of tissue-mimicking material (TMM) based on the expression of the measurement of uncertainty. Six 60-mm-diameter TMMs were fabricated, three 10 mm thick and three 20 mm thick. The experimental setup comprised two ultrasonic transducers, acting as transmitter or receiver depending on the stage of the measurement protocol, both with a nominal center frequency of 5 MHz and an element diameter of 12.7 mm. A sine burst of 20 cycles and 20-V peak-to-peak amplitude at 5 MHz excited the transmitter transducer, producing a maximum pressure of 0.06 MPa. The measurement method was based on the through-transmission substitution immersion technique. The speed of sound measurement system was validated using a calibrated stainless-steel cylinder as reference material, and normalized errors were <0.8. The attenuation coefficient measurement method was validated using replicated measurements under repeatability conditions. The normalized error between the two measurement sets was <1. The proposed uncertainty models for the measurements of the speed of sound and the attenuation coefficient can help other laboratories develop their own uncertainty models. These validated measurement methods can be used to certify a TMM as a reference material for biotechnological applications.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ultrassonografia / Imagens de Fantasmas / Modelos Biológicos Tipo de estudo: Diagnostic_studies / Prognostic_studies Idioma: En Revista: Ultrasound Med Biol Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Brasil País de publicação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ultrassonografia / Imagens de Fantasmas / Modelos Biológicos Tipo de estudo: Diagnostic_studies / Prognostic_studies Idioma: En Revista: Ultrasound Med Biol Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Brasil País de publicação: Reino Unido