Study of neutron scattering contribution on Hp(10) and H*(10) calibration in the Brazilian National Low Scattering Laboratory.
Freitas, B M; Pereira, W W; Patrão, K C S; Fonseca, E S; Mauricio, C L P.
Afiliação
  • Freitas BM; Instituto de Radioproteção e Dosimetria, Av. Salvador Allende, s/n, Recreio, Rio de Janeiro, CEP: 22783-127 RJ, Brazil.
  • Pereira WW; Instituto de Radioproteção e Dosimetria, Av. Salvador Allende, s/n, Recreio, Rio de Janeiro, CEP: 22783-127 RJ, Brazil.
  • Patrão KC; Instituto de Radioproteção e Dosimetria, Av. Salvador Allende, s/n, Recreio, Rio de Janeiro, CEP: 22783-127 RJ, Brazil.
  • Fonseca ES; Instituto de Radioproteção e Dosimetria, Av. Salvador Allende, s/n, Recreio, Rio de Janeiro, CEP: 22783-127 RJ, Brazil.
  • Mauricio CL; Instituto de Radioproteção e Dosimetria, Av. Salvador Allende, s/n, Recreio, Rio de Janeiro, CEP: 22783-127 RJ, Brazil claudia@ird.gov.br.
Radiat Prot Dosimetry ; 161(1-4): 176-80, 2014 Oct.
Article em En | MEDLINE | ID: mdl-24984874
The neutron scattering at the Low Scattering Laboratory of the Brazilian National Neutron Laboratory has been studied using three different methods. The measurements have been done with a traceable standard (241)Am-Be from source-to-detector distances of 0.52-3.00 m. The obtained results with the variation distance methods are in agreement. Measurements with a large shadow cone are not worth for larger distances due to overshadowing. As the quantity required in a calibration is the response of the device being calibrated to the scattered neutron component in order to subtract this from the total response, for these purposes, the distance variation method must be used for each device. To quantify absolutely the scattering contribution on the quantity rates of fluence, Hp(10) and H*(10) in irradiation procedures, a Bonner sphere spectrometer with the shadow cone was employed. The evaluated scattering correction factor value may be employed for a distance of 1.00 m.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Radiometria / Espalhamento de Radiação / Nêutrons Tipo de estudo: Prognostic_studies País/Região como assunto: America do sul / Brasil Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Brasil País de publicação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Radiometria / Espalhamento de Radiação / Nêutrons Tipo de estudo: Prognostic_studies País/Região como assunto: America do sul / Brasil Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Brasil País de publicação: Reino Unido