Stokes polarimetry using analysis of the nonlinear voltage-retardance relationship for liquid-crystal variable retarders.
Rev Sci Instrum
; 85(3): 033104, 2014 Mar.
Article
em En
| MEDLINE
| ID: mdl-24689561
We present a method for using liquid-crystal variable retarders (LCVR's) with continually varying voltage to measure the Stokes vector of a light beam. The LCVR's are usually employed with fixed retardance values due to the nonlinear voltage-retardance behavior that they show. The nonlinear voltage-retardance relationship is first measured and then a linear fit of the known retardance terms to the detected signal is performed. We use known waveplates (half-wave and quarter-wave) as devices to provide controlled polarization states to the Stokes polarimeter and we use the measured Stokes parameters as functions of the orientation of the axes of the waveplates as an indication of the quality of the polarimeter. Results are compared to a Fourier analysis method that does not take into account the nonlinear voltage-retardance relationship and also to a Fourier analysis method that uses experimental voltage values to give a linear retardance function with time. Also, we present results of simulations for comparison.