Analysis of stress induced by screws in the vertebral fixation system.
Acta Ortop Bras
; 22(1): 17-20, 2014.
Article
em En
| MEDLINE
| ID: mdl-24644414
OBJECTIVE: To compare, using photoelasticity, internal stress produced by USS II type screw with 5.2 and 6.2 mm external diameters, when submitted to three different pullout strengths. METHODS: Two photoelastic models were especially made. The simulation was performed using loads of 1.8, 2.4 e 3.3 kgf.The fringe orders were evaluated around the screws. In all the models analyzed the shear stress were calculated. RESULTS: Independently of the applied load, the smaller screw showed higher values of shear stress. CONCLUSION: According to the analysis performed, we observed that the place of highest stress was in the first thread of the lead, close to the head of the screws. Experimental study.