Elimination of the copper-zinc interference in anodic stripping voltammetry by addition of a complexing agent.
Brandes Marques, A L; Chierice, G O.
Afiliação
  • Brandes Marques AL; Dep. Tecnologia Quimica/UFMA, Caixa Postal 369, CEP 13560, Sao Carlos/SP, Brazil.
Talanta ; 38(7): 735-9, 1991 Jul.
Article em En | MEDLINE | ID: mdl-18965213
The addition of cyanide and Triton X-100 suppress the formation of the CuZn intermetallic compound in ASV making it possible to determine traces of zinc(II) in the presence of an excess of copper ions. The precision of the method (5%) and the accuracy (error of 1.4%) in sea water are satisfactory.