Your browser doesn't support javascript.
loading
In vitro measurement accuracy of an electronic apex locator in teeth with simulated apical root resorption.
Goldberg, Fernando; De Silvio, Ana C; Manfré, Susana; Nastri, Natalia.
Afiliação
  • Goldberg F; Department of Endodontics, School of Dentistry, USAL-AOA, Buenos Aires, Argentina.
J Endod ; 28(6): 461-3, 2002 Jun.
Article em En | MEDLINE | ID: mdl-12067130
The purpose of this in vitro study was to evaluate the accuracy of Root ZX apex locator to determine the working length in teeth with simulated apical root resorption. Fifty extracted, single-rooted, human teeth with mature apices were used in this study. An irregular cavity defect was drilled at the apex of each tooth simulating an apical root resorption. Three operators used the Root ZX to measure the working lengths, comparing the electronic readings with the direct visual measurements. The Root ZX was 62.7%, 94.0%, and 100.0% accurate to within 0.5 mm, 1 mm, and 1.5 mm of the direct visual measurements, respectively. Statistically significant differences were observed between operator A and B and A and C (p < 0.01), but no significant differences were detected between operator B and C (p > 0.01).
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Reabsorção da Raiz / Raiz Dentária / Cavidade Pulpar / Odontometria Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: J Endod Ano de publicação: 2002 Tipo de documento: Article País de afiliação: Argentina País de publicação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Reabsorção da Raiz / Raiz Dentária / Cavidade Pulpar / Odontometria Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: J Endod Ano de publicação: 2002 Tipo de documento: Article País de afiliação: Argentina País de publicação: Estados Unidos