[Evaluation of commercial equipment used in blood banks in Córdoba for the detection of anti-HTLV-I/II antibodies]. / Evaluación de los equipos comerciales utilizados en los bancos de sangre de Córdoba para la detección de anticuerpos anti HTLV-I/II.
Rev Argent Microbiol
; 32(4): 202-5, 2000.
Article
em Es
| MEDLINE
| ID: mdl-11149153
In order to assess the efficiency of currently used screening tests, Abbott HTLV-I/HTLV-II EIA, Vironostika HTLV-I/II Organon Teknika, Particle Agglutination (PA) assay Serodia Fujirebio Inc. (Tokyo, Japan) for HTLV-I/II antibody detection in blood donors samples, a panel of 100 sera from different blood banks of Córdoba city were studied. An "in house" indirect immunofluorescence assay (IFA) was used as reference test. The correlation rates were: 66% for Abbott HTLV-I/HTLV-II EIA, 97% for Vironostika HTLV-I/II Organon Teknika EIA and 99% for PA Serodia. Vironostika HTLV-I/II Organon Teknika EIA and PA Serodia assay proved to be more reliable for HTlV-I/II antibody screening in blood donors from Córdoba, yielding a very low rate of false positive results as compared with Abbot HTLV-I/HTLV-II EIA.