Comparação da umidade e temperatura dos gases inalados com alto ou baixo fluxo de gases frescos e com ou sem permutador de calor e umidade em anestesia pediátrica / Comparison of the humidity of inhaled gases with low or high fresh gasflow and with or without a heat and moisture exchanger in pediatric patients
Botucatu; s.n; 2013. 91 p. ilus, tab.
Thesis em Pt | LILACS | ID: lil-751051
Biblioteca responsável: BR33.1
RESUMO
O valor mínimo de umidade absoluta (UA) mais frequentemente considerado para prevenção dos efeitos deletérios de gases secos sobre o epitélio das vias aéreas durante anestesia é de 20 mg H2O/L. Como as crianças possuem menor volume minuto, a umidificação dos gases no sistema ventilatório circular pode ser menor na criança que no adulto. O permutador de calor e umidade (PCU) é dispositivo utilizado para umidificar e aquecer adicionalmente os gases inalados durante a ventilação artificial. Com o objetivo de avaliar as propriedades de umidificação de sistema ventilatório circular durante anestesia pediátrica, realizou-se estudo para comparar a temperatura e umidade dos gases inalados com baixo ou alto fluxo de gases frescos (FGF) e com ou sem PCU no circuito ventilatório. Quarenta crianças foram alocadas aleatoriamente em um de quatro grupos segundo a ventilação pulmonar com sistema circular com baixo (1L/min) ou alto (3L/min) FGF sem PCU (grupos 1L e 3L) ou com PCU (grupos PCU1L e PCU3L). A temperatura e UA dos gases inalados foram mensuradas durante 80 minutos após a conexão do paciente ao circuito ventilatório. As médias da temperatura do gás inalado foram maiores nos grupos que utilizaram PCU (PCU1L 30,3 ± 1,1°C; PCU3L 29,3± 1,2oC) quando comparadas às dos grupos que não utilizaram PCU (1L 27,0 ± 1,2°C; 3L 27,1 ± 1,5oC) (p < 0,0001). As médias de UA dos gases inalados foram maiores nos grupos com PCU em relação aos grupos sem PCU e maiores nos grupos de baixo FGF que nos grupos de alto FGF [(PCU1L 25 ± 1 mg H20/L) > (PCU3L 23 ± 2 mg H20/L) > (1L 17 ± 1 mg H20/L) > (3L 14 ± 1 mg H20/L)] (p<0,0001)...
ABSTRACT
An inhaled gas absolute humidity (AH) of 20 mg H2O.L-1 is the minimum value most considered as the threshold necessary for preventing the deleterious effects of dry gas on the epithelium of the airways during anesthesia. Because children have small minute ventilation, we hypothesized that the humidification of a circle breathing system is lower in children compared with adults. A heat and moisture exchanger (HME) is a device that can be used to further humidify inhaled gases during anesthesia. In order to evaluate the humidifying properties of a circle breathing system during pediatric anesthesia we compared the temperature and humidity of inhaled gases under low or high fresh gas flow (FGF) conditions and with or without an HME in the breathing circuit. Forty children were randomly allocated into one of four groups according to the ventilation of their lungs by a circle breathing system with low (1L.min-1) or high (3L.min-1) FGF without an HME (1L and 3L groups) or with an HME (HME1L and HME3L groups). The temperature and AH of inhaled gases were measured for 80 minutes after connecting the patient to the breathing circuit. The mean inhaled gas temperature was higher in HME groups (HME1L 30.3 ± 1.1°C; HME3L 29.3 ± 1.2°C) compared with no-HME groups (1L 27.0 ± 1.2°C; 3L 27.1 ± 1.5°C) (P < 0.0001). The mean inhaled gas AH was higher in HME than no-HME groups, and higher in low-flow than high-flow groups [(HME1L 25 ± 1 mg H2O.L-1) > (HME3L 23 ± 2 mg H2O.L-1) > (1L 17 ± 1 mg H2O.L-1) > (3L 14 ± 1 mg H2O.L-1)] (P < 0.0001). In a pediatric circle breathing system, the use of neither high nor low FGF provides the minimum humidity level of the inhaled gases thought to reduce the risk of dehydration of airways...
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: LILACS Assunto principal: Anestesia com Circuito Fechado / Anestesia por Inalação Tipo de estudo: Clinical_trials Limite: Child / Female / Humans / Male Idioma: Pt Ano de publicação: 2013 Tipo de documento: Thesis País de publicação: Brasil
Buscar no Google
Coleções: 01-internacional Base de dados: LILACS Assunto principal: Anestesia com Circuito Fechado / Anestesia por Inalação Tipo de estudo: Clinical_trials Limite: Child / Female / Humans / Male Idioma: Pt Ano de publicação: 2013 Tipo de documento: Thesis País de publicação: Brasil