Evaluación de la prueba ICT Malaria P. f/P.v ( AMRAD ® ) para la detección de P. falciparum y P. vivax en una zona endémica de la Amazonía peruana / ICT malaria P.f./P.v. (AMRAD ®) test for detecting P. falciparum and P. vivax in an endemic area in peruvian amazon jungle
Rev. peru. med. exp. salud publica
; 19(1): 39-42, ene.-mar. 2002. mapas, tab
Article
em Es
| LILACS, LIPECS, INS-PERU
| ID: lil-498587
Biblioteca responsável:
PE14.1
RESUMEN
Objetivo:
Evaluar el rendimiento del ICT MALARIA P.f/P.v (AMRAD®) en pacientes febriles del departamento de Loreto, Perú. Materiales ymétodos:
estudio transversal realizado de agosto a setiembre del 2000, en pacientes con historia de fiebre (temperatura axilar 37,5°C) en los últimos 3 días, sin un foco aparente. Se obtuvo una muestra de sangre para gota gruesa cuya lectura se realizó por un microscopista experto. Simultáneamente, se realizó la prueba rápida por un asistente de campo. Se calculó la sensibilidad, especificidad, valores predictivos, exactitud e índice de concordancia utilizando la gota gruesa como prueba de oro.Resultados:
se incluyeron 79 pacientes de 5 meses - 70 años de edad. La sensibilidad y especificidad del AMRAD ® fueron 60,0 y 70,0 por ciento, respectivamente. El valor predictivo positivo (VPP), valor predictivo negativo (VPN) y exactitud de la prueba fueron 36,0 por ciento, 88,8 por ciento y 70,9 por ciento, respectivamente. Se encontró pobre concordancia de la prueba con la gota gruesa (kappa=0,279).Conclusión:
es necesario continuar con la evaluación de las pruebas rápidas para un diagnóstico oportuno y un tratamiento efectivo de la malaria.ABSTRACT
Objective:
To assess the ICT MALARIA P.f/P.v (AMRAD®) in febrile patients in Loreto Department, Peru. Material andmethods:
A cross-sectional study performed in febrile (axillary temperature 37.5°C) patients with no apparent cause for the fever, from August to September 2000. A blood sample for performing thick smear examination was obtained. Well-trained personnel performed this test. Also, a field assistant performed the rapid test. Sensitivity, specificity, predictive values, accurateness and concordance index were calculated using the thick blood smear as the gold standard.Results:
79 patients were included (5 months - 70 years of age). Sensitivity and specificity for the AMRAD® were 60,0 and 70,0 per cent, respectively. Positive predictive value (PPV), negative predictive value (NPV) and accuracy of the test were respectively 36,0, 88,8 and 70,9 per cent. The test had poor concordance with the thick smear test (kappa=0,279).Conclusion:
It is necessary to continue assessing rapid testing methods in order to have a timely diagnosis and early therapy for malaria.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
INS-PERU
/
LILACS
/
LIPECS
Assunto principal:
Sensibilidade e Especificidade
/
Técnicas de Laboratório Clínico
/
Malária
Tipo de estudo:
Diagnostic_studies
/
Observational_studies
/
Prevalence_studies
/
Prognostic_studies
/
Risk_factors_studies
Limite:
Adolescent
/
Adult
/
Female
/
Humans
/
Male
Idioma:
Es
Revista:
Rev. peru. med. exp. salud publica
Assunto da revista:
SAUDE PUBLICA
Ano de publicação:
2002
Tipo de documento:
Article
País de afiliação:
Peru
País de publicação:
Peru