Análise química e estrutural de componente femoral de prótese total do quadril removido após fratura por fadiga: relato de caso / Chemical and structural analysis of total hip prosthesis femoral component retrieved after fatigue fracture: a case report
Rev. bras. ortop
; 37(6): 256-258, jun. 2002. ilus, tab
Article
em Pt
| LILACS
| ID: lil-322141
Biblioteca responsável:
BR26.1
ABSTRACT
The authors present one case of stem fatigue fracture ofa cemented total hip prosthesis, which took place 43 months after its implantation. There was good cement thickness, adequate stem positioning, and absence of radiographic signs of loosening at the revision. The retrieved implant and an unused stem of the same type (control) were analyzed by macrographic inspection, metallographic and chemical analysis. The implants were found to have improper attributes of manufacturing and structural nature
Buscar no Google
Coleções:
01-internacional
Base de dados:
LILACS
Assunto principal:
Seguimentos
/
Fraturas do Quadril
Tipo de estudo:
Observational_studies
/
Prognostic_studies
Limite:
Aged
/
Humans
/
Male
Idioma:
Pt
Revista:
Rev. bras. ortop
Assunto da revista:
ORTOPEDIA
Ano de publicação:
2002
Tipo de documento:
Article
País de afiliação:
Brasil
País de publicação: