Sensibilidad y especificidad de pruebas inmunocromatográficas utilizadas en el nuevo algoritmo de diagnóstico de VIH en Bolivia / Sensitivity and specificity of inmunocromatographic tests used in the new algorithm of HIV diagnosis in Bolivia
Gac. méd. boliv
; 41(2): 6-8, Dec. 2018. ilus., tab.
Article
em Es
| LILACS, LIBOCS
| ID: biblio-988155
Biblioteca responsável:
BO138.1
RESUMEN
OBJETIVO:
evaluar la sensibilidad y especificidad de dos pruebas rápidas utilizadas en el nuevo algoritmo de diagnóstico de VIH en Bolivia, Alere Determine TM HIV 1/2 como prueba de tamizaje y Uni-Gold TM HIV como prueba confirmatoria.MÉTODOS:
estudio descriptivo, no experimental. Se utilizaron 60 muestras de suero provenientes de diferentes establecimientos de salud de Cochabamba con resultados reactivos para VIH, enviadas a LABIMED desde enero a junio de 2016 para confirmación. Las 60 muestras (27 positivas y 33 negativas para VIH) fueron procesadas con Alere Determine TM HIV 1/2 y Uni-Gold TM HIV.RESULTADOS:
alere Determine TM HIV 1/2 presentó una sensibilidad del 100% y una especificidad del 54,5%. Uni-Gold TM HIV reportó una especificad del 97% y una sensibilidad del 92,6%Conclusiones:
la sensibilidad de Alere Determine TM HIV 1/2 cumplió con el criterio del Ministerio de Salud y Deportes de Bolivia (Prueba rápida de tamizaje debe tener una sensibilidad ≥ 99,8%). La especificidad de Uni-Gold TM HIV en este estudio no alcanzó el criterio de especificidad del Ministerio (Prueba rápida de confirmación debe tener una especificidad ≥ 99,9%).ABSTRACT
OBJECTIVE:
to evaluate the sensitivity and specificity of the two rapid tests used in the new algorithm of HIV diagnosis in Bolivia, Alere DetermineTM HIV 1/2 as a screening test and Uni-Gold TM HIV as a confirmatory test.METHODS:
this is a descriptive and non-experimental study. Sixty serum samples were used with reactive results for HIV from different health establishments in Cochabamba sent to LABIMED from January to June 2016 for HIV confirmation. The 60 samples (27 positive and 33 negative for HIV) were tested with Alere DetermineTM HIV 1/2 and Uni-GoldTM HIV.RESULTS:
Alere DetermineTM HIV 1/2 presented a sensitivity of 100 % and a specificity of 54,5%. Uni-GoldTM HIV reported a sensitivity of 92,6% and a specificity of 97%.Conclusions:
the sensitivity of Alere Determine TM HIV 1/2 fulfilled the criteria of the Ministry of Health and Sports of Bolivia (rapid screening test must have a sensitivity ≥ 99.8%). The specificity of Uni-GoldTM HIV in this study did not fulfill the Ministry's specificity criterion (rapid confirmation test must have a specificity ≥99.9%).Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
LIBOCS
/
LILACS
Assunto principal:
Anticorpos Anti-HIV
Tipo de estudo:
Diagnostic_studies
/
Prognostic_studies
/
Screening_studies
País/Região como assunto:
America do sul
/
Bolivia
Idioma:
Es
Revista:
Gac. méd. boliv
Assunto da revista:
MEDICINA
Ano de publicação:
2018
Tipo de documento:
Article
País de afiliação:
Bolívia
País de publicação:
Bolívia