El RCP Lázaro como alternativa al perfeccionamiento de la reanimación cardiopulmonar
Cardoso Totoy, Darwin Rodrigo; Riera Sampedro, Valeria Alexandra; Díaz Armas, María Teresa; Solis Cartas, Urbano.
Afiliação
  • Cardoso Totoy, Darwin Rodrigo; s.af
  • Riera Sampedro, Valeria Alexandra; s.af
  • Díaz Armas, María Teresa; s.af
  • Solis Cartas, Urbano; s.af
Rev. cuba. reumatol ; 25(2)jun. 2023.
Article em Es | LILACS-Express | LILACS | ID: biblio-1565522
Biblioteca responsável: CU1.1
RESUMEN

Introducción:

el paro cardiorrespiratorio de la situación más peligrosa que puede enfrentar un ser humano; para su recuperación se necesita actuar inmediatamente y de forma correcta. El tiempo de comienzo de la reanimación cardiopulmonar es vital para lograr la supervivencia del paciente.

Objetivo:

socializar las ventajas que ofrece el RCP Lázaro en la asistencia de urgencia a pacientes con parada cardiorrespiratoria.

Métodos:

se realizó una investigación aplicada, con diseño experimental que parte de la identificación de una necesidad de salud y se procede a la ideación, diseño y creación de un prototipo de equipo de reanimación cardiopulmonar asistida denominado RCP Lázaro. Se confeccionó un equipo multidisciplinario que permitió obtener los planos mecánicos, ergonómicos y eléctricos que permiten un funcionamiento adecuado bajo el mando de un software diseñado específicamente para el equipo. Después de terminada la fase de modelación y diseño se realizaron pruebas funcionales que finalmente fueron superadas.

Resultado:

El RCP Lázaro es un prototipo superior en varios sentidos a los modelos previos existentes en el mercado. Presenta un diseño ergonómico que permite un mejor ajuste a la complexión física de los pacientes; puede ser utilizado en niños y adultos; presenta posibilidad de acople de desfibrilador; presenta 4 ciclos de compresiones/respiraciones; reajustable la altura del pistón para garantizar adaptación previa a las compresiones torácicas; batería de 6 horas de duración, entendible en modo ahorro de energía hasta 8 horas y facilidades para ensamblaje, transporte y acoplamiento al paciente.

Conclusiones:

El RCP Lázaro paso de forma satisfactoria las pruebas de validación, mostrando parámetros de funcionamiento adecuado para la función que está diseñada; corresponde entonces proceder a la prueba de campo antes de iniciar su producción en serie.
ABSTRACT

Introduction:

cardiorespiratory arrest is the most dangerous situation that a human being can face; for your recovery you need to act immediately and correctly. The start time of cardiopulmonary resuscitation is vital to achieve patient survival.

Objective:

to socialize the advantages offered by Lazarus CPR in emergency care for patients with cardiorespiratory arrest.

Methods:

an applied investigation was carried out, with an experimental design that starts from the identification of a health need and proceeds to the ideation, design and creation of a prototype of assisted cardiopulmonary resuscitation equipment called Lazarus CPR. A multidisciplinary team was created that allowed obtaining the mechanical, ergonomic and electrical plans that allow proper operation under the command of software specifically designed for the equipment. After the modeling and design phase was completed, functional tests were carried out that were finally passed.

Result:

The Lazarus CPR is a superior prototype in several ways to the previous models on the market. It presents an ergonomic design that allows a better adjustment to the physical complexion of the patients; it can be used in children and adults; presents the possibility of defibrillator coupling; presents 4 cycles of compressions/breaths; the height of the piston can be readjusted to guarantee adaptation prior to chest compressions; 6-hour battery life, understandable in energy saving mode up to 8 hours and facilities for assembly, transport and coupling to the patient.

Conclusions:

The Lazarus CPR satisfactorily passed the validation tests, showing adequate operating parameters for the function that is designed; It is then appropriate to proceed to the field test before starting its production in series.
Palavras-chave
Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Idioma: Es Revista: Rev. cuba. reumatol Assunto da revista: MEDICINA Ano de publicação: 2023 Tipo de documento: Article País de publicação: Cuba
Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Idioma: Es Revista: Rev. cuba. reumatol Assunto da revista: MEDICINA Ano de publicação: 2023 Tipo de documento: Article País de publicação: Cuba