Validación del método analítico por cromatografía líquida de alta resolución para la cuantificación de conservantes en jugo de naranja / Validation of the analytical method by liquid chromatography of high resolution for the quantification of preservatives in orange juice
Con-ciencia (La Paz)
; 8(2): 61-74, 2020. ilus., tab.
Article
em Es
| LILACS, LIBOCS
| ID: biblio-1147992
Biblioteca responsável:
BO138.1
RESUMEN
INTRODUCCIÓN:
la validación es el establecimiento de la evidencia documental de que un procedimiento analítico conducirá, con un alto grado de seguridad a la obtención de resultados precisos y exactos dentro de las especificaciones y los atributos de calidad previamente establecidos.OBJETIVO:
Validar el método analítico por cromatografía líquida de alta resolución para la cuantificación de conservantes en jugo de naranja.MÉTODOS:
el método fue validado siguiendo los parámetros referidos en la USP 40 NF 35 especificidad, linealidad, precisión (repetibilidad instrumental, repetibilidad del método y precisión intermedia) y exactitud.RESULTADOS:
el método analítico propuesto es selectivo/específico, presenta una excelente linealidad dentro del rango de 100 a 1000 ppm, la Precisión presentó resultados de CV < 2% siendo óptimos. La recuperación se encuentra entre 97 a 102% demostrando excelente exactitud.CONCLUSIÓN:
se cumplieron todos los parámetros y criterios de validación por lo tanto el método es específico, selectivo, lineal en el intervalo de concentraciones de 100 a 1000 ppm, sensible, preciso, reproducible y exactoABSTRACT
INTRODUCTION:
validation is the establishment of documentary evidence that an analytical procedure will lead, with a high degree of security, to obtaining precise and exact results within the specifications and previously established quality attributes.OBJECTIVE:
validate the analytical method by high resolution liquid chromatography for the quantification of preservatives in orange juice.METHODS:
the method was validated following the parameters referred to in USP 40 - NF 35, they are specificity, linearity, precision (instrumental repeatability, repeatability of the method and intermediate precision) and accuracy.RESULTS:
the proposed analytical method is selective/specific, it presents an excellent linearity within the range of 100 to 1000 ppm, the precision presented results of CV <2% being optimal. The recovery is between 97 to 102%CONCLUSION:
all the parameters and validation criteria were met, therefore the method is specific, selective, linear in the concentration range of 100 to 1000 ppm, sensitive, precise, reproducible and exactPalavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
LIBOCS
/
LILACS
Assunto principal:
Cromatografia Líquida
/
Citrus sinensis
Tipo de estudo:
Diagnostic_studies
Idioma:
Es
Revista:
Con-ciencia (La Paz)
Assunto da revista:
Farmacologia
/
Teraputica
/
Toxicologia
Ano de publicação:
2020
Tipo de documento:
Article
País de afiliação:
Bolívia
País de publicação: