Determination of Fe, Co, Mn and Ni in Synthetic Diamonds by Inductively Coupled Plasma Atomic Emission Spectrometry / 分析化学
Article en Zh | WPRIM | ID: wpr-671730
Biblioteca responsable: WPRO
ABSTRACT
A method for the determination of Fe, Co, Mn and Ni in synthetic diamonds by inductively coupled plasma atomic emission spectrometry ( ICP-AES) was proposed. The synthetic diamond sample was decomposed completely, while the sample was burned in air at 1000 ℃ for 10 h, and then a mixed acid of H2 SO4 , aqua regia and HClO4 was used for the dissolving the residue of the sample. In this method, the limits of detection of Fe, Co, Mn and Ni were 0. 0147, 0. 0018, 0. 0006 and 0. 0027 mg/L, respectively. Under the optimum condition, Fe, Co, Mn and Ni in synthetic diamond sample were determined. The values of RSDs (n=7) were less than 0. 5%. The recoveries of added standard were 94. 0%-105. 0%.
Palabras clave
Texto completo: 1 Base de datos: WPRIM Idioma: Zh Revista: Chinese Journal of Analytical Chemistry Año: 2014 Tipo del documento: Article