Your browser doesn't support javascript.
loading
Establishment and validation of finger-ring dosimeter standard calibration quality / 中国医学装备
China Medical Equipment ; (12): 10-12,13, 2016.
Article en Zh | WPRIM | ID: wpr-603931
Biblioteca responsable: WPRO
ABSTRACT
Objective:According to the technical standard of ISO and current regulations in China, the X/γ standard calibration quality for the finger-ring dosimeter irradiation was established in Beijing SSDL, it can ensure the traceability of surveillance data within countrywide.Methods: Air kerma-rate was determined in the X-ray beams,137Cs and60Co by a standard dosimeter, combined transfer coefficientHp(0.07) provided by standards, and confirmed condition of X/γ standard calibration quality for the finger-ring dosimeter irradiation. Using finger-ring TLD dosimeter verified the radiation beam qualities Hp on the ISO finger-phantom.Results: The linearity and energy response of the TLD dosimeter met ISO and national standard in Beijing SSDL.Conclusion: The X/γ standard radiation field can be used for the calibration of finger-ring dosimeter standard curves and experiment of research, and can further promote the comparability and reliability of the surveillance data.
Palabras clave
Texto completo: 1 Base de datos: WPRIM Tipo de estudio: Prognostic_studies Idioma: Zh Revista: China Medical Equipment Año: 2016 Tipo del documento: Article
Texto completo: 1 Base de datos: WPRIM Tipo de estudio: Prognostic_studies Idioma: Zh Revista: China Medical Equipment Año: 2016 Tipo del documento: Article