Your browser doesn't support javascript.
loading
The study on the application of active fixation electrodes in patients with right ventricular outflow tract septal pacing / 中国医师进修杂志
Article en Zh | WPRIM | ID: wpr-394610
Biblioteca responsable: WPRO
ABSTRACT
Objective To explore the feasibility of right ventricular outflow tract (RVOT) septal pacing. Methods Fifty-two patients implanted with pacemaker were randomly divided into two groups. One group underwent the RVOT septal pacing with active fixation electrodes (active fixation electrodes group, 22 cases),and the other group underwent the right ventricular apical pacing with passive fixation electrodes (passive fixation electrodes group, 30 cases). The parameters of two groups in operation were recorded. Results The successful rate of RVOT septal pacing was 100%, active fixation electrodes group took more time to plant electrodes and expose in X-ray than passive fixation electrodes group [ (56.13±1.18)min vs (15.42±1.24) min, (18.77±6.14) min vs (10.12±8.14) min, P<0.05]. The width of the QRS wave of active fixation electrodes group was narrower than that of passive fixation electrodes group [(0.14±0.02)ms vs (0.16±0.04) ms, P< 0.05]. The parameters of two groups were stable in follow-up period and none of electrodes was extracted. Conclusion The usage of active fixation electrodes in patients with RVOT septal pacing is feasible and safe.
Palabras clave
Texto completo: 1 Base de datos: WPRIM Idioma: Zh Revista: Chinese Journal of Postgraduates of Medicine Año: 2009 Tipo del documento: Article
Texto completo: 1 Base de datos: WPRIM Idioma: Zh Revista: Chinese Journal of Postgraduates of Medicine Año: 2009 Tipo del documento: Article