This paper presents the preliminary design of data acquisition system of a portable uroflowmeter. The system uses double-hole cantilever pressure sensor. The signal is transferred to ATmega644PA microprogrammed control unit (MCU), converted by A/D (analog to digital) convertor. Then the further data are processed and get the corresponding relationship of weight-time and two curves of urine flow and urinary flow rate. In the measurement accuracy of the device about urine flow, two factors about the placement and height of the data acquisition are analyzed to show the accuracy of the equipment through the Origin 8.0 data analysis software. The design is characterized by low cost and high speed of data collection, real-time, high accuracy.