Your browser doesn't support javascript.
loading
Postoperative Life-Threatening Recurrent Ventricular Arrhythmia Triggered by the Swan-Ganz Catheter in a Patient Undergoing Off-Pump Coronary Artery Bypass Surgery
Article en En | WPRIM | ID: wpr-109979
Biblioteca responsable: WPRO
ABSTRACT
Recurrent ventricular arrhythmia can be fatal and cause serious complications, particularly when it is caused immediately after an operation. Incorrect placement of a Swan-Ganz catheter can trigger life-threatening ventricular arrhythmia, but even intensive care specialists tend to miss this fact. Here, we report a case of recurrent ventricular arrhythmia causing a severe hemodynamic compromise; the arrhythmia was induced by a severely angulated Swan-Ganz catheter. The recurrent ventricular arrhythmia was not controlled by any measures including repositioning of the catheter, until the complete removal of the Swan-Ganz catheter. It is necessary to keep in mind that the position of the pulmonary artery catheter should be promptly checked if there is intractable recurrent ventricular arrhythmia.
Asunto(s)
Palabras clave
Texto completo: 1 Base de datos: WPRIM Asunto principal: Arritmias Cardíacas / Arteria Pulmonar / Especialización / Cuidados Críticos / Puente de Arteria Coronaria Off-Pump / Catéteres / Hemodinámica Límite: Humans Idioma: En Revista: The Korean Journal of Thoracic and Cardiovascular Surgery Año: 2014 Tipo del documento: Article
Texto completo: 1 Base de datos: WPRIM Asunto principal: Arritmias Cardíacas / Arteria Pulmonar / Especialización / Cuidados Críticos / Puente de Arteria Coronaria Off-Pump / Catéteres / Hemodinámica Límite: Humans Idioma: En Revista: The Korean Journal of Thoracic and Cardiovascular Surgery Año: 2014 Tipo del documento: Article