Your browser doesn't support javascript.
loading
Comparative analysis of domestic Octoparms and imported Celect inferior vena cava filter in interventional treatment of venous thromboembolism / 实用放射学杂志
Journal of Practical Radiology ; (12): 456-459,463, 2024.
Article en Zh | WPRIM | ID: wpr-1020237
Biblioteca responsable: WPRO
ABSTRACT
Objective To compare and analyze the application value of domestic Octoparms and imported Celect inferior vena cava filter(IVCF)in the interventional treatment of venous thromboembolism(VTE).Methods Forty patients with VTE were randomly divided into Octoparms group(experimental group)and Celect group(control group)according to the double-blinded method of the central random system.All the patients underwent filter placement,catheter-directed thrombolysis and filter retrieval.The primary end point was the success of filter placement and retrieval,and the secondary end point included indwelling complications such as the occurrence of pulmonary embolism(PE)and filter tilt and migration.Results Forty patients were enrolled in this study,22 patients and 18 patients were divided into the experimental group and the control group,respectively.Among them,11 cases were identified with right lower extremity deep vein thrombosis,29 cases with left lower extremity deep vein thrombosis,17 cases with PE,and 6 cases with inferior vena cava thrombosis.The success rate of IVCF placement was 100%in all participants.Immediately after filter place-ment,the angle of filter tilt was(3.8±2.3)° in the experimental group and(4.9±2.8)° in the control group(t=1.44,P=0.16).Filter retrieval was successful in 21 cases(21/22,95.5%)of the experimental group and 17 cases(17/18,95.5%)of the control group.There was no significant difference between the two groups(t=0.14,P=0.89).The mean indwelling time of filter was(8.0±2.1)days in the experimental group and(9.7±3.1)days in the control group(t=0.73,P=0.47).The angle of filter tilt was(5.3±3.4)° in the experimental group and(5.7±7.7)° in the control group(t=0.19,P=0.85).There was no significant difference for filter placement and retrieval between the two groups(t=0.48 and 2.00,P=0.06 and 0.64,respectively).There were no complications of filter migration,strut penetration or new PE in both groups.Conclusion The application value of domestic Octoparms and impor-ted Celect IVCF is similar in interventional treatment of VTE.
Palabras clave
Texto completo: 1 Base de datos: WPRIM Idioma: Zh Revista: Journal of Practical Radiology Año: 2024 Tipo del documento: Article
Texto completo: 1 Base de datos: WPRIM Idioma: Zh Revista: Journal of Practical Radiology Año: 2024 Tipo del documento: Article