A universal method for geometric correction of magnetic resonance images for stereotactic neurosurgery.
Stereotact Funct Neurosurg
; 66(1-3): 137-40, 1996.
Article
en En
| MEDLINE
| ID: mdl-8938945
Accurate stereotactic navigation depends strongly upon the spatial fidelity of the image used for registration. Clinically significant levels of geometric distortion are present in standard MR images, limiting their utility. A technique for correction of all geometric distortions in spine echo MR images was assessed in a prospective clinical trial of 19 stereotactic craniotomies. The Euclidean error in target registration between CT and MR was significantly reduced, from 3.833 +/- 0.992 to 1.986 +/- 0.605 mm. The results of this clinical trial support the incorporation of this MR image rectification protocol into standard clinical practice.