An unopened titanium Greenfield IVC filter: intravascular ultrasound to reveal associated thrombus and aid in filter opening.
Goertzen, T C; McCowan, T C; Garvin, K L; Lieberman, R P; LeVeen, R F.
Afiliación
  • Goertzen TC; Department of Radiology, University of Nebraska Medical Center, Omaha 68198.
Cardiovasc Intervent Radiol ; 16(4): 251-3, 1993.
Article en En | MEDLINE | ID: mdl-8402791
A titanium Greenfield filter did not open following placement in the infrarenal inferior vena cava (IVC). Abdominal radiograph and cavogram showed no definite reason for filter malfunction. Intravascular ultrasound (IVUS) demonstrated the unopened filter in the IVC with thrombus binding the legs. The thrombus was disrupted with a catheter, and the filter completely expanded with a balloon. IVUS documented full-filter opening in addition to residual thrombus in the filter following manipulation.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Trombosis / Titanio / Vena Cava Inferior / Filtros de Vena Cava / Ultrasonografía Intervencional Tipo de estudio: Risk_factors_studies Límite: Aged / Female / Humans Idioma: En Revista: Cardiovasc Intervent Radiol Año: 1993 Tipo del documento: Article Pais de publicación: Estados Unidos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Trombosis / Titanio / Vena Cava Inferior / Filtros de Vena Cava / Ultrasonografía Intervencional Tipo de estudio: Risk_factors_studies Límite: Aged / Female / Humans Idioma: En Revista: Cardiovasc Intervent Radiol Año: 1993 Tipo del documento: Article Pais de publicación: Estados Unidos