Your browser doesn't support javascript.
loading
EORTC radiotherapy group quality assurance: mechanical checks and beam alignments of megavoltage equipment.
Van Dam, J; Johansson, K A; Bridier, A; Sernbo, G; Hansson, U.
Afiliación
  • Van Dam J; University Hospital St. Rafaël, Leuven, Belgium.
Radiother Oncol ; 29(2): 91-6, 1993 Nov.
Article en En | MEDLINE | ID: mdl-8310159
In 1987 mechanical checks of megavoltage units and simulators were included in the on-site physics program of the EORTC. The results reported were obtained in 16 different centres and concern 23 accelerators, 14 cobalt units and 14 simulators. In general, the deviations observed for accelerators and simulators are smaller than for cobalt units. A single score, based on the deviations observed for the mechanical checks, is attributed to each centre.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Radioterapia de Alta Energía / Neoplasias Tipo de estudio: Prognostic_studies Límite: Humans País/Región como asunto: Europa Idioma: En Revista: Radiother Oncol Año: 1993 Tipo del documento: Article País de afiliación: Bélgica Pais de publicación: Irlanda
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Radioterapia de Alta Energía / Neoplasias Tipo de estudio: Prognostic_studies Límite: Humans País/Región como asunto: Europa Idioma: En Revista: Radiother Oncol Año: 1993 Tipo del documento: Article País de afiliación: Bélgica Pais de publicación: Irlanda