Disintegration of Multiple Parts of a Bone Transport Nail During Planned Removal: A Case Report.
Halvorson, Ryan; Kandemir, Utku.
Afiliación
  • Halvorson R; Department of Orthopaedic Surgery, University of California, San Francisco, California.
JBJS Case Connect ; 14(3)2024 Jul 01.
Article en En | MEDLINE | ID: mdl-39270038
ABSTRACT
CASE A 34-year-old man with a distal tibia bone defect was treated with an intramedullary bone transport nail (Precice Bone Transport System, NuVasive). During planned removal after successful treatment, 7 separate subcomponents of the nail became disconnected and had to be separately removed using specialized instrumentation. This occurred despite adherence to the manufacturer's recommended technique for nail removal and in the absence of clinical or radiographic evidence of implant failure.

CONCLUSION:

When planning for implant removal, surgeons should be aware of potential intraoperative disconnection of subcomponents of this magnetic bone transport nail and ensure that equipment for retrieval (e.g., very long endoscopy forceps) is available.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fracturas de la Tibia / Clavos Ortopédicos / Remoción de Dispositivos Límite: Adult / Humans / Male Idioma: En Revista: JBJS Case Connect Año: 2024 Tipo del documento: Article Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fracturas de la Tibia / Clavos Ortopédicos / Remoción de Dispositivos Límite: Adult / Humans / Male Idioma: En Revista: JBJS Case Connect Año: 2024 Tipo del documento: Article Pais de publicación: Estados Unidos