Recording maximal intercuspation and border positions of the mandible with intraoral scanner using the acquisition software's multi-occlusion function.
J Adv Prosthodont
; 16(4): 221-230, 2024 Aug.
Article
en En
| MEDLINE
| ID: mdl-39221415
ABSTRACT
PURPOSE:
This in vitro study was conducted to investigate the accuracy of intraoral scanner (IOS) for recording maximal intercuspal position (MIP) and border positions of the mandible. MATERIALS ANDMETHODS:
Maxillary and mandibular master casts were articulated in MIP, protrusive, and lateral interocclusal position sequentially on a semi-adjustable articulator. For each articulation relation, sites of occlusal contacts (SOCs) and sites of clearance (SCs) were identified on the master casts with articulating paper (reference sites). IOS was used to take full arch scans and nine virtual interocclusal records (VIRs) for virtual articulation of models. Virtual SOCs and SCs were detected with 3D processing software and compared to those identified with the articulating paper. Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were calculated for each articulation relation.RESULTS:
For MIP, IOS showed adequate sensitivity and NPV of 100%, and specificity and a PPV of 99%. For protrusive position, the IOS showed a sensitivity and a NPV of 100%, a high PPV of 86%, and a specificity of 83%. For lateral positions, the specificity and the PPV were high (93% and 79%, respectively), but the sensitivity and the NPV were below the clinically acceptable limits (28% and 56%, respectively).CONCLUSION:
IOS displayed clinically acceptable accuracy for recording MIP and protrusive border mandibular position. However, IOS had less accuracy for lateral border mandibular position.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Idioma:
En
Revista:
J Adv Prosthodont
Año:
2024
Tipo del documento:
Article
País de afiliación:
Egipto
Pais de publicación:
Corea del Sur