In Vitro Assessment of a New Block Design for Implant Crowns with Functional Gradient Fabricated with Resin Composite and Zirconia Insert.
Gutiérrez Robledo, Nicolás; Punset Fuste, Miquel; Rodríguez-Contreras, Alejandra; García Marro, Fernando; Manero Planella, José María; Figueras-Álvarez, Oscar; Roig Cayón, Miguel.
Afiliación
  • Gutiérrez Robledo N; School of Dentistry, Universitat Internacional de Catalunya (UIC), 08195 Barcelona, Spain.
  • Punset Fuste M; Independent Research, 28028 Madrid, Spain.
  • Rodríguez-Contreras A; Biomaterials, Biomechanics and Tissue Engineering Group (BBT), Universitat Politècnica de Catalunya (UPC), 08019 Barcelona, Spain.
  • García Marro F; UPC Innovation and Technology Center (CIT-UPC), Universitat Politècnica de Catalunya (UPC), 08034 Barcelona, Spain.
  • Manero Planella JM; Barcelona Research Centre in Multiscale Science and Engineering, Universitat Politècnica de Catalunya (UPC), 08019 Barcelona, Spain.
  • Figueras-Álvarez O; Reserach Institute San Joan de Déu (IRSJD), 08950 Barcelona, Spain.
  • Roig Cayón M; Biomaterials, Biomechanics and Tissue Engineering Group (BBT), Universitat Politècnica de Catalunya (UPC), 08019 Barcelona, Spain.
Materials (Basel) ; 17(15)2024 Aug 02.
Article en En | MEDLINE | ID: mdl-39124479
ABSTRACT
This study aims to evaluate and compare the mechanical resistance, fatigue behavior and fracture behavior of different CAD/CAM materials for implant crowns. Eighty-eight implant crowns cemented-screwed with four sample groups two monolithic G1 Zirconia (control) and G3 composite and two bi-layered G2 customized zirconia/composite and G4 prefabricated zirconia/composite. All static and dynamic mechanical tests were conducted at 37 °C under wet conditions. The fractographic evaluation of deformed and/or fractured samples was evaluated via electron microscopy. Statistical analysis was conducted using Wallis tests, which were performed depending on the variables, with a confidence interval of 95%, (p < 0.05). The Maximum Fracture Strength values displayed by the four groups of samples showed no statistically significant differences. The crown-abutment material combination influenced the failure mode of the restoration, transitioning from a fatigue fracture type located at the abutment-analog connection for monolithic materials (G1 and G3) to a brittle fracture located in the crown for bi-layered materials (G2 and G4). The use of layered crown materials with functional gradients appears to protect the crown/abutment connection area by partially absorbing the applied mechanical loads. This prevents catastrophic mechanical failures, avoiding long chairside time to solve these kinds of complications.