Robotic MR-guided high dose rate brachytherapy needle implantation in the prostate (ROBiNSon)-a proof-of-concept study.
Rasing, Marnix J A; van Schelven, Leonard J; van Lier, Astrid L H M W; Boskovic, Ernest; van Son, Marieke; Schokker, Rogier I; van Vliet-van den Ende, Katelijne M; Kelder, Daniëlle D; Verkerk, Jacqueline; van de Pol, Sandrine M G; Lagendijk, Jan J W; Peters, Max; van der Voort van Zyp, Jochem R N; Moerland, Marinus A.
Afiliación
  • Rasing MJA; Department of Radiation Oncology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • van Schelven LJ; Department of Medical Technology and Clinical Physics, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • van Lier ALHMW; Department of Radiation Oncology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • Boskovic E; Department of Medical Technology and Clinical Physics, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • van Son M; University General Practice, Amsterdam University Medical Center, De Boelelaan 1117, 1007 MB Amsterdam, The Netherlands.
  • Schokker RI; Department of Radiation Oncology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • van Vliet-van den Ende KM; Department of Radiation Oncology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • Kelder DD; Department of Radiation Oncology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • Verkerk J; Department of Radiation Oncology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • van de Pol SMG; Department of Radiation Oncology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • Lagendijk JJW; Department of Radiation Oncology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • Peters M; Department of Radiation Oncology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • van der Voort van Zyp JRN; Department of Radiation Oncology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
  • Moerland MA; Department of Radiation Oncology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
Phys Med Biol ; 69(17)2024 Aug 21.
Article en En | MEDLINE | ID: mdl-39084657
ABSTRACT
Objective.A robotic needle implant device for MR-guided high-dose-rate (HDR) prostate brachytherapy was developed. This study aimed to assess the feasibility and spatial accuracy of HDR brachytherapy using the robotic device, for a single intraprostatic target point.Approach.Five patients were treated from November 2019-June 2022 with the robot. The robot fits a 1.5 T MR scanner and the needle can be shifted and angulated. An intraprocedural MR scan was fused with the diagnostic MR and one preplanned needle position was selected for robotic insertion. The needle entry point and angles were set for a needle tip target point within the intraprostatic target volume. The needle was tapped stepwise towards the target point pneumatically. Final needle position was verified with MR, followed by plan optimization and dose delivery. Any remaining planned needles were inserted manually. Needle tip to geometrical target error (NTG-error) was defined as the deviation of the actual tip position relative to the predefined geometric target point, using MR-coordinates. Needle tip to treatment target error (NTT-error) was defined as the deviation of the actual tip position relative to the treatment target point, using fused MR-images pre- and post-needle implantation taking into account prostate deformation. Difference between NTT-error and NTG-error and fiducial marker shifts indicated prostate movement. For determining prostate deformation, the Jaccard index and prostate volumes were assessed.Main results.The robotic device was able to tap the needle to the planned depth for all patients. Mean robotic procedure duration was 142 min. NTG-error was 3.2 (range 1.1-6.7) mm and NTT-error 4.5 (range 2.6-9.6) mm. Marker displacements were smaller than 3 mm. No treatment-related acute toxicity was reported. Feasibility of needle placement within the prostate was considered adequate.Significance.MR-guided robotic needle insertion is feasible with a mean geometric accuracy of 3.2 mm and <3 mm prostate movement.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Neoplasias de la Próstata / Dosificación Radioterapéutica / Robótica / Braquiterapia / Imagen por Resonancia Magnética / Radioterapia Guiada por Imagen / Agujas Límite: Humans / Male Idioma: En Revista: Phys Med Biol Año: 2024 Tipo del documento: Article País de afiliación: Países Bajos Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Neoplasias de la Próstata / Dosificación Radioterapéutica / Robótica / Braquiterapia / Imagen por Resonancia Magnética / Radioterapia Guiada por Imagen / Agujas Límite: Humans / Male Idioma: En Revista: Phys Med Biol Año: 2024 Tipo del documento: Article País de afiliación: Países Bajos Pais de publicación: Reino Unido