Late recognition of cardiac implantable electronic device misplacement in left ventricle: a case report.
Bahrami, Fatemeh; Sattartabar, Babak; Larti, Farnoosh; Rahmanian, Mehrzad; Mollazadeh, Reza.
Afiliación
  • Bahrami F; Department of Cardiology, School of Medicine, Imam Khomeini Hospital Complex, Tehran University of Medical Sciences, PO Box: 1419733141, Tehran, Iran.
  • Sattartabar B; Department of Cardiology, Research Department, Tehran Heart Center, Tehran University of Medical Sciences, Tehran, Iran.
  • Larti F; Department of Cardiology, School of Medicine, Imam Khomeini Hospital Complex, Tehran University of Medical Sciences, PO Box: 1419733141, Tehran, Iran.
  • Rahmanian M; Department of Cardiothoracic Surgery, School of Medicine, Imam Khomeini Hospital Complex, Tehran University of Medical Sciences, PO Box: 1419733141, Tehran, Iran.
  • Mollazadeh R; Department of Cardiology, School of Medicine, Imam Khomeini Hospital Complex, Tehran University of Medical Sciences, PO Box: 1419733141, Tehran, Iran.
Future Cardiol ; 20(5-6): 269-274, 2024 Apr 25.
Article en En | MEDLINE | ID: mdl-38899509
ABSTRACT

Background:

Cardiac electronic device implantation may be associated with complications. Case

Summary:

This is a report of inadvertent implantation of implantable cardioverter-defibrillator lead through an unrecognized sinus venosus atrial septal defect into the left ventricle that was not diagnosed early after implantation. Six months later chest x-ray showed an abnormal lead course that was confirmed with echocardiography as to be in the left ventricle. Surgical removal of the implantable cardioverter-defibrillator lead, repair of atrial septal defect, and correction of abnormal pulmonary venous connections were performed. Meanwhile, follow-up of the patient receiving a new dual chamber permanent pacemaker from the contralateral side and discussion of the aforementioned complication are addressed.

Conclusion:

Early diagnosis of device implantation complication is of paramount importance and prevents potential catastrophic complications.
This is a report of a middle-aged woman who received a heart battery (pacemaker) without proper initial indication and faced unnecessary consequences/complications. The wire was supposed to be implanted in the right side of the heart, but due to the congenital heart disease of the patient (that was not diagnosed), the wire was erroneously passed through the congenital defect to the left side of the heart. After the diagnosis of the problem, we referred her to a heart surgeon to correct the congenital heart defect and also pull the wire out. After open surgery, a new pacemaker (with two wires) was implanted in the patient without any additional complications.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Desfibriladores Implantables / Ventrículos Cardíacos Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Future Cardiol Asunto de la revista: CARDIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Irán Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Desfibriladores Implantables / Ventrículos Cardíacos Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Future Cardiol Asunto de la revista: CARDIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Irán Pais de publicación: Reino Unido