Reliability of Ultrasound Measurements of the Median Nerve in Asymptomatic Subjects Using a Handheld Device.
Alfuraih, Abdulrahman M; Aldahlawi, Rana Hussain; Habib, Yomna S; Alhowimel, Ahmed S; Bedewi, Mohamed Abdelmohsen.
Afiliación
  • Alfuraih AM; Radiology and Medical Imaging Department, College of Applied Medical Sciences, Prince Sattam bin Abdulaziz University, Kharj 16278, Saudi Arabia.
  • Aldahlawi RH; Department of Radiological Sciences, College of Applied Medical Sciences, King Saud University, Riyadh 14511, Saudi Arabia.
  • Habib YS; Department of Radiology, University Hospital, Prince Sattam bin Abdulaziz University, Kharj 16278, Saudi Arabia.
  • Alhowimel AS; Department of Health and Rehabilitation Sciences, College of Applied Medical Sciences, Prince Sattam bin Abdulaziz University, Kharj 16278, Saudi Arabia.
  • Bedewi MA; Department of Internal Medicine, College of Medicine, Prince Sattam bin Abdulaziz University, Kharj 16278, Saudi Arabia.
Sensors (Basel) ; 24(11)2024 May 27.
Article en En | MEDLINE | ID: mdl-38894235
ABSTRACT
This study investigated the reliability of measuring the median nerve cross-sectional area (CSA) at the carpal tunnel inlet using a handheld ultrasound device (HUD) compared to a standard ultrasound system, focusing on intra- and inter-operator reproducibility among novice and expert operators. Employing a prospective cross-sectional design, 37 asymptomatic adults were assessed using both devices, with measurements taken by an expert with over five years of experience and a novice with less than six months. The CSA was determined using manual tracing and ellipse methods, with reproducibility evaluated through intraclass correlation coefficients (ICCs) and agreement assessed via Bland-Altman plots. Results showed a high degree of agreement between the devices, with excellent intra-operator reproducibility (ICC > 0.80) for the expert, and moderate reproducibility for the novice (ICCs ranging from 0.539 to 0.841). Inter-operator reliability was generally moderate, indicating acceptable consistency across different experience levels. The study concludes that HUDs are comparable to standard ultrasound systems for assessing median nerve CSA in asymptomatic subjects, with both devices providing reliable measurements. This supports the use of HUDs in diverse clinical environments, particularly where access to traditional ultrasound is limited. Further research with a larger sample and symptomatic patients is recommended to validate these findings.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ultrasonografía / Nervio Mediano Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Sensors (Basel) Año: 2024 Tipo del documento: Article País de afiliación: Arabia Saudita Pais de publicación: Suiza

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ultrasonografía / Nervio Mediano Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Sensors (Basel) Año: 2024 Tipo del documento: Article País de afiliación: Arabia Saudita Pais de publicación: Suiza