Standardizing technical parameters and terms for abdominopelvic photon-counting CT: laying the groundwork for innovation and evidence sharing.
Leng, Shuai; Toia, Giuseppe V; Hoodeshenas, Safa; Ramirez-Giraldo, Juan Carlos; Yagil, Yoad; Maltz, Jonathan S; Boedeker, Kirsten; Li, Ke; Baffour, Francis; Fletcher, Joel G.
Afiliación
  • Leng S; Department of Radiology, Mayo Clinic, 200 First Street SW, Rochester, MN, 55905, USA.
  • Toia GV; Departments of Radiology and Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, WI, USA.
  • Hoodeshenas S; Department of Radiology, Mayo Clinic, 200 First Street SW, Rochester, MN, 55905, USA.
  • Ramirez-Giraldo JC; Siemens Medical Solutions USA Inc, Malvern, PA, USA.
  • Yagil Y; PD CT/AMI R&D Advanced Development, Philips Medical Systems, Haifa, Israel.
  • Maltz JS; Molecular Imaging and Computed Tomography, GE Healthcare, Waukesha, WI, USA.
  • Boedeker K; Canon Medical Systems Corporation, Otawara, Japan.
  • Li K; Departments of Radiology and Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, WI, USA.
  • Baffour F; Department of Radiology, Mayo Clinic, 200 First Street SW, Rochester, MN, 55905, USA.
  • Fletcher JG; Department of Radiology, Mayo Clinic, 200 First Street SW, Rochester, MN, 55905, USA. Fletcher.Joel@mayo.edu.
Abdom Radiol (NY) ; 49(9): 3261-3273, 2024 Sep.
Article en En | MEDLINE | ID: mdl-38769199
ABSTRACT
Photon-counting detector CT (PCD-CT) is a new technology that has multiple diagnostic benefits including increased spatial resolution, iodine signal, and radiation dose efficiency, as well as multi-energy imaging capability, but which also has unique challenges in abdominal imaging. The purpose of this work is to summarize key features, technical parameters, and terms, which are common amongst current abdominopelvic PCD-CT systems and to propose standardized terminology (where none exists). In addition, user-selectable protocol parameters are highlighted to facilitate both scientific evaluation and early clinical adoption. Unique features of PCD-CT systems include photon-counting detectors themselves, energy thresholds and bins, and tube potential considerations for preserved spectral separation. Key parameters for describing different PCD-CT systems are reviewed and explained. While PCD-CT can generate multi-energy images like dual-energy CT, there are new types of images such as threshold images, energy bin images, and special spectral images. The standardized terms and concepts herein build upon prior interdisciplinary consensus and have been endorsed by the newly created Society of Abdominal Radiology Photon-counting CT Emerging Technology Commission.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Radiografía Abdominal / Tomografía Computarizada por Rayos X / Fotones Límite: Humans Idioma: En Revista: Abdom Radiol (NY) Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Radiografía Abdominal / Tomografía Computarizada por Rayos X / Fotones Límite: Humans Idioma: En Revista: Abdom Radiol (NY) Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos