Assessment of average glandular dose in mammography practice of a teaching hospital in Ghana.
Kyei, Kofi Adesi; Anim-Sampong, Samuel; Ahulu, Eugene Nartey; Antwi, William Kwadwo; Daniels, Joseph.
Afiliación
  • Kyei KA; Department of Radiography, University of Ghana, Korle Bu, Accra, Ghana.
  • Anim-Sampong S; National Radiotherapy Oncology and Nuclear Medicine Centre, Korle Bu Teaching Hospital, Accra, Ghana.
  • Ahulu EN; Department of Radiography, University of Ghana, Korle Bu, Accra, Ghana.
  • Antwi WK; Department of Radiography, University of Ghana, Korle Bu, Accra, Ghana.
  • Daniels J; Department of Radiography, University of Ghana, Korle Bu, Accra, Ghana.
Pan Afr Med J ; 47: 42, 2024.
Article en En | MEDLINE | ID: mdl-38681097
ABSTRACT

Introduction:

above the age of 40, women are advised to begin breast examinations and screenings for early detection of breast cancer. The average glandular dose (AGD) provides dosimetric information about the quantity of radiation received by the mammary glands during mammographic exposures. There is, therefore, the need to analyse the radiation dose received by patients presenting for mammography examinations.

Methods:

a retrospective cross-sectional design was carried out on the data of 663 participants, conveniently sampled between the months of July 2021 and June 2022. Paired T-test was used to compare imaging parameters for cranio-caudal (CC), medio-lateral (ML), automatic exposure control (AEC), manual exposure control (MEC), and left and right breast. Pearson´s correlation was used to test for relationship between imaging parameters and AGD.

Results:

the mean AGD per exposure was 1.9 ± 0.7 mGy for CC projections and 2.3 ± 1.2 mGy for ML projections. The mean AGD per examination for the study was 4.1 ± 1.4 mGy. A positive correlation was found between AGD per examination and exposure factors (tube loading and tube voltage), compressed breast thickness, and compression force. Patient age had no statistically significant relationship with the AGD per examination.

Conclusion:

average glandular dose (AGD) was consistent with other findings in literature studies. It was also observed that MEC yielded lower AGD per exposure values than AEC. There was no significant difference in the mean AGD per exposure for left and right breasts.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosis de Radiación / Neoplasias de la Mama / Mamografía / Hospitales de Enseñanza Límite: Adult / Aged / Female / Humans / Middle aged País/Región como asunto: Africa Idioma: En Revista: Pan Afr Med J Año: 2024 Tipo del documento: Article País de afiliación: Ghana Pais de publicación: Uganda

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosis de Radiación / Neoplasias de la Mama / Mamografía / Hospitales de Enseñanza Límite: Adult / Aged / Female / Humans / Middle aged País/Región como asunto: Africa Idioma: En Revista: Pan Afr Med J Año: 2024 Tipo del documento: Article País de afiliación: Ghana Pais de publicación: Uganda