Your browser doesn't support javascript.
loading
Comparison of Intraocular Lens Power Prediction Accuracy Between 2 Swept-Source Optical Coherence Tomography Biometry Devices.
Sorkin, Nir; Zadok, Rotem; Savini, Giacomo; Kan-Tor, Yoav; Benjamini, Yuval; Levinger, Eliya; Bardugo, Judith; Abulafia, Adi.
Afiliación
  • Sorkin N; Department of Ophthalmology, Tel Aviv Medical Center, Tel Aviv, Israel; Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel. Electronic address: nir.sorkin@gmail.com.
  • Zadok R; Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.
  • Savini G; IRCCS Bietti Foundation, Rome, Italy.
  • Kan-Tor Y; The Alexander Grass Center for Bioengineering, School of Computer Science and Engineering, Hebrew University of Jerusalem, Jerusalem, Israel.
  • Benjamini Y; Department of Statistics and Data Science, The Hebrew University of Jerusalem, Jerusalem, Israel.
  • Levinger E; Department of Ophthalmology, Tel Aviv Medical Center, Tel Aviv, Israel; Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.
  • Bardugo J; Department of Ophthalmology, Shaare Zedek Medical Center and the Hebrew University-Hadassah Medical School, Jerusalem, Israel.
  • Abulafia A; Department of Ophthalmology, Shaare Zedek Medical Center and the Hebrew University-Hadassah Medical School, Jerusalem, Israel.
Am J Ophthalmol ; 265: 156-164, 2024 Sep.
Article en En | MEDLINE | ID: mdl-38643892
ABSTRACT

PURPOSE:

To compare intraocular lens (IOL) power prediction accuracy of the Eyestar 900 (EyeS900) and the IOLMaster 700 (IOLM700) based on estimated and measured posterior corneal power.

DESIGN:

Retrospective, interinstrument reliability study.

METHODS:

Setting:

Institutional.

PARTICIPANTS:

Two hundred twenty-five eyes of 225 cataract surgery patients. MEASUREMENTS Patients underwent measurements by both devices preoperatively. MAIN OUTCOME

MEASURES:

Spherical Equivalent Prediction Error (SEQ-PE), spread of the SEQ-PE (precision) and the absolute SEQ-PE (accuracy) of each device using Barrett Universal II (BUII) formula with either estimated posterior keratometry (E-PK) or measured posterior keratometry (M-PK).

RESULTS:

Trimmed mean SEQ-PEs of EyeS900 E-PK, EyeS900 M-PK, IOLM700 E-PK, and IOLM700 M-PK were 0.03, 0.08, 0.02, and 0.09 D, respectively with no significant differences between EyeS900 E-PK and IOLM700 E-PK (P = 0.31) as well as between EyeS900 M-PK and IOLM700 M-PK (P = 0.31). Statistically significant SEQ-PE differences were found when E-PK and M-PK were compared, regardless of the device used, showing hyperopic SEQ-PE in M-PK calculations. Excellent correlation and agreement in SEQ-PE were found between the devices for both E-PK (P < 0.001, r = 0.848, mean bias +0.01 D, 95% LOA of -0.32 to +0.34 D) and M-PK (P < 0.001, r = 0.776, mean bias -0.01 D, 95% LOA of -0.42 to +0.39 D). No significant differences were found comparing absolute SEQ-PE and precision of the devices.

CONCLUSION:

The Eyestar 900 and the IOLMaster 700 show comparable IOL power prediction accuracy by the BUII formula using either estimated or measured posterior keratometry. An adjusted lens factor may be required for BUII when utilizing measured posterior keratometry in both devices.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Refracción Ocular / Biometría / Implantación de Lentes Intraoculares / Tomografía de Coherencia Óptica / Óptica y Fotónica / Lentes Intraoculares Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Am J Ophthalmol Año: 2024 Tipo del documento: Article Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Refracción Ocular / Biometría / Implantación de Lentes Intraoculares / Tomografía de Coherencia Óptica / Óptica y Fotónica / Lentes Intraoculares Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Am J Ophthalmol Año: 2024 Tipo del documento: Article Pais de publicación: Estados Unidos