Measurement proprieties of the CROM instrument for assessing head posture, neck retraction and protraction.
Elizagaray-García, Ignacio; Obispo-Villamayor, Álvaro; Prats-Martínez, Celia; Prieto-Hernández, Gonzalo; Carvalho, Gabriela Ferreira.
Afiliación
  • Elizagaray-García I; CranioSPain Research Group, Centro Superior de Estudios Universitarios La Salle, 28023, Madrid, Spain. Electronic address: jeliga@lasallecampus.es.
  • Obispo-Villamayor Á; Department/Group, Centro Superior de Estudios Universitarios La Salle, Universidad Autónoma de Madrid, 28023, Madrid, Spain.
  • Prats-Martínez C; Department/Group, Centro Superior de Estudios Universitarios La Salle, Universidad Autónoma de Madrid, 28023, Madrid, Spain.
  • Prieto-Hernández G; Department/Group, Centro Superior de Estudios Universitarios La Salle, Universidad Autónoma de Madrid, 28023, Madrid, Spain.
  • Carvalho GF; Department of Physiotherapy, Faculty of Health, Safety and Society, Furtwangen University, Furtwangen, Germany.
Musculoskelet Sci Pract ; 71: 102950, 2024 06.
Article en En | MEDLINE | ID: mdl-38574577
ABSTRACT

BACKGROUND:

The CROM instrument is widely used clinically and in research to measure neck range of motion. However, its measurement proprieties during the assessment of protraction and retraction movements were not examined so far.

OBJECTIVE:

To analyse the intra- and inter-rater reliability, the concurrent validity of the CROM for measuring head posture, retraction and protraction in healthy subjects.

METHODS:

Thirty-three asymptomatic subjects were recruited and assigned in a random order to one of two raters. After a 10-min break, they were examined by a second rater (Assessment 1). After a 30-min break, both raters repeated the examination (Assessment 2). The examination consisted of measuring the head posture, maximum head protraction and maximum retraction. Each movement was repeated 3 times and measured simultaneously with the CROM and with a 3D capture system laboratory.

RESULTS:

The intra-rater reliability of the CROM was excellent for both raters for head posture and all head movements (ICC>0.9, 95% CI 0.82-0.99, p < 0.01). The inter-rater reliability was excellent for head posture (ICC>0.95, 95% CI 0.92-0.98, p < 0.01) and good-to-excellent for all movements at both time-points (ICC = 0.73-0.98, 95%CI 0.45-0.99, p < 0.01). The validity analysis showed moderate-to-strong correlation between instruments for the head posture and head movements [(r) = -0.47 to -0.78), 95% CI 0.99 to -0.24, p < 0.01].

CONCLUSION:

The CROM instrument has good-to-excellent reliability and adequate validity for measuring cervical position and displacement in the sagittal plane.