Very-early detection of atrial fibrillation after ablation evaluated by a wearable ECG-patch predicts late blanking period recurrence: Preliminary data from a prospective registry.
Int J Cardiol Heart Vasc
; 51: 101369, 2024 Apr.
Article
en En
| MEDLINE
| ID: mdl-38420510
ABSTRACT
Introduction:
Atrial fibrillation (AF) ablation represents a safe and effective procedure to restore sinus rhythm. The idea that post-procedural AF episodes - during the blanking period - are not considered treatment failure has been increasingly challenged. The E-Patch, a single-use adhesive electrode, facilitates extended continuous ECG monitoring for 120 h. This pilot study aims to assess the effectiveness of this ambulatory monitoring device and investigate whether very-early AF recurrence correlates with delayed blanking period ablation outcomes.Methods:
We conducted a single-center, prospective, longitudinal study, including consecutive post-ablation patients monitored with the E-patch. The ability of the device to continuously record was analyzed, as well as the occurrence of AF episodes during external 7-day loop-recorder in the 2nd-month post-ablation.Results:
We included 40 patients, median age 62 years (IQR 56-70). E-Patch monitoring was obtained for a median of 118 h (IQR 112-120), with no discomfort nor interpretation artefacts. Very-early AF recurrence was detected in 11 (27.5 %) patients, with a median AF burden of 7 % (IQR 6 %-33 %). Late-blanking period AF was detected in 13 (33 %) of the external 7-day loop recordings. Of the 11 patients that had very-early AF recurrence, 10 (91 %) had late-blanking AF. Very-early AF detection showed 77 % (95 % CI 64 %-90 %) sensitivity and 96 % (95 % CI 90-100 %) specificity in predicting late-blanking AF, with a non-parametric ROC curve AUC of 0.903 (95 % 0.797--1.0).Conclusion:
The E-Patch was able to detect very-early AF during an extended period. Very-early AF detection emerges as a predictor of AF recurrence during the late blanking period post-ablation.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Idioma:
En
Revista:
Int J Cardiol Heart Vasc
Año:
2024
Tipo del documento:
Article
País de afiliación:
Portugal
Pais de publicación:
Irlanda